Multi-dimensional respiratory motion tracking from markerless optical surface imaging based on deformable mesh registration
Authors:
- Aurora FassiAFAurora Fassi
- Marco RiboldiMRMarco Riboldi
- Pietro CerveriPCPietro Cerveri
- Guido BaroniGBGuido Baroni
- David SarrutDSDavid Sarrut
- 14 December 2011
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 57 (2) , 357-373
Abstract
Real-time optical surface imaging systems offer a non-invasive way to monitor intra-fraction motion of a patient's thorax surface during radiotherapy treatments. Due to lack of point correspondence in dynamic surface acquisition, such systems cannot currently provide 3D motion tracking at specific surface landmarks, as available in optical technologies based on passive markers. We propose to apply deformable mesh registration to extract surface point trajectories from markerless optical imaging, thus yielding multi-dimensional breathing traces. The investigated approach is based on a non-rigid extension of the iterative closest point algorithm, using a locally affine regularization. The accuracy in tracking breathing motion was quantified in a group of healthy volunteers, by pair-wise registering the thoraco-abdominal surfaces acquired at three different respiratory phases using a clinically available optical system. The motion tracking accuracy proved to be maximal in the abdominal region, where breathing motion mostly occurs, with average errors of 1.09 mm. The results demonstrate the feasibility of recovering multi-dimensional breathing motion from markerless optical surface acquisitions by using the implemented deformable registration algorithm. The approach can potentially improve respiratory motion management in radiation therapy, including motion artefact reduction or tumour motion compensation by means of internal/external correlation models.Keywords
This publication has 18 references indexed in Scilit:
- Geometric accuracy of a novel gimbals based radiation therapy tumor tracking systemRadiotherapy and Oncology, 2011
- A multiple points method for 4D CT image sortingMedical Physics, 2011
- Clinical Accuracy of the Respiratory Tumor Tracking System of the CyberKnife: Assessment by Analysis of Log FilesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- 4D laser camera for accurate patient positioning, collision avoidance, image fusion and adaptive approaches during diagnostic and therapeutic proceduresMedical Physics, 2008
- 3D optoelectronic analysis of interfractional patient setup variability in frameless extracranial stereotactic radiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- A phantom evaluation of a stereo‐vision surface imaging system for radiotherapy patient setupMedical Physics, 2005
- The space of human body shapesACM Transactions on Graphics, 2003
- Evaluation of respiratory movement during gated radiotherapy using film and electronic portal imagingInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2002
- Rigid, affine and locally affine registration of free-form surfacesInternational Journal of Computer Vision, 1996
- A method for registration of 3-D shapesIEEE Transactions on Pattern Analysis and Machine Intelligence, 1992