Abstract
An analytical method was developed to quantify the aromatic monomer products of lignin derived reductive catalytic fractionation (RCF) by ultra-high pressure liquid chromatography with diode array detection (UHPLC-DAD). Chromatographic separation of aromatic analytes was achieved using a reverse-phase C18 column with a mobile phase gradient. Analytes quantified include guaiacyl, syringyl, and p-hydroxyphenyl monomers as well as hydroxycinnamic acid derivatives. This protocol utilizes an Agilent Technologies 1290 Infinity UHPLC-DAD system as referenced in the Materials section. A similar chromatography and detection system may be used, however some parameter nomenclature may differ depending on the manufacturer.