Pressure and stress transients in autoinjector devices
- 6 August 2018
- journal article
- research article
- Published by Springer Nature in Drug Delivery and Translational Research
- Vol. 8 (5) , 1238-1253
Abstract
Spring-actuated autoinjectors delivering viscous drug solutions resulting from large drug concentrations require large spring forces which can create high peak pressures and stresses within syringes. The high peak pressures and stresses can lead to device failure. Measurements with a suite of novel instrumentation and analysis using numerical simulation explain the peak pressures and peak stresses as originating from mechanical impacts between moving components, the large acceleration of the components, and surprisingly, the production of tension waves in the liquid resulting in cavitation. The presence and intensity of cavitation depend on relative timing between the pressurization and the acceleration of the syringe, which, in turn, depend on the size and location of an air gap inside the syringe. We show that production of localized but very high pressures can result from shock wave focusing in the conical section of the syringe.Keywords
Funding Information
- Amgen
This publication has 9 references indexed in Scilit:
- New autoinjector technology for the delivery of subcutaneous methotrexate in the treatment of rheumatoid arthritisExpert Review of Medical Devices, 2014
- Pen and Autoinjector Drug Delivery Devices2013
- Self-Injection Devices2013
- Advances in parenteral injection devices and aids2012
- Applied Mechanics of Solids2009
- Challenges in the development of high protein concentration formulationsJournal of Pharmaceutical Sciences, 2004
- Cavitation And Bubble Dynamics1995
- Wave Propagation in Solids and Fluids1988
- Supersonic Flow and Shock Waves1976