Development of a high-Performance MALDI-TOF mass spectrometer utilizing a spiral ion trajectory
Authors:
- 1 July 2007
- journal article
- Published by American Chemical Society (ACS) in Journal of the American Society for Mass Spectrometry
- Vol. 18 (7) , 1318-1323
Abstract
A novel MALDI-TOF mass spectrometer that utilizes a spiral ion trajectory was developed. In this mass spectrometer, the ions sequentially passed through four toroidal electrostatic sectors and revolved along a figure-eight-shaped orbit on a particular projection plane. Each toroidal electrostatic sector had eight stories, and during multiple revolutions, the ion trajectory shifted perpendicular to the projection plane in every cycle, thereby generating a spiral trajectory. The flight path length of one cycle was 2.1 m; therefore, when the ions completed eight cycles, the total flight path length was 17 m. By adopting an ion optical system that had a flight path length five times longer than that in the commonly used reflectron ion optical system, the mass dependence on the mass resolving power was reduced, while improving the mass accuracy of the mass measurements. The basic performance of the system was tested by using standard peptides or the tryptic digest of bovine serum albumin. A mass resolving power of 80,000 (full width at half maximum) was achieved at m/z = 2564 (ACTH18-39). An improved mass accuracy less than 2 ppm was realized over a wide m/z range of 500 to 3000 by correction using one or two internal standard substances.Keywords
This publication has 12 references indexed in Scilit:
- The design and characteristic features of a new time-of-flight mass spectrometer with a spiral ion trajectoryJournal of the American Society for Mass Spectrometry, 2005
- Construction of a new multi-turn time-of-flight mass spectrometerJournal of Mass Spectrometry, 2000
- Perfect space and time focusing ion optics for multiturn time of flight mass spectrometersInternational Journal of Mass Spectrometry, 2000
- A new multi-passage time-of-flight mass spectrometer at JAISTNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1999
- On the initial velocity of ions generated by matrix-assisted laser desorption ionization and its effect on the calibration of delayed extraction time-of-flight mass spectraJournal of the American Society for Mass Spectrometry, 1997
- Mass Resolution Improvement by Incorporation of Pulsed Ion Extraction in a Matrix-Assisted Laser Desorption/Ionization Linear Time-of-Flight Mass SpectrometerAnalytical Chemistry, 1995
- Delayed extraction matrix‐assisted laser desorption time‐of‐flight mass spectrometryRapid Communications in Mass Spectrometry, 1995
- Laser desorption ionization of proteins with molecular masses exceeding 10,000 daltonsAnalytical Chemistry, 1988
- Multiple-focusing time-of-flight mass spectrometers Part II. TOFMS with equal energy accelerationInternational Journal of Mass Spectrometry and Ion Physics, 1972
- An Ion ``Velocitron''Review of Scientific Instruments, 1948