Learning-based hypothesis fusion for robust catheter tracking in 2D X-ray fluoroscopy
Authors:
- 1 June 2011
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 1097-1104
Abstract
Catheter tracking has become more and more important in recent interventional applications. It provides real time navigation for the physicians and can be used to control a motion compensated fluoro overlay reference image for other means of guidance, e.g. involving a 3D anatomical model. Tracking the coronary sinus (CS) catheter is effective to compensate respiratory and cardiac motion for 3D overlay navigation to assist positioning the ablation catheter in Atrial Fibrillation (Afib) treatments. During interventions, the CS catheter performs rapid motion and non-rigid deformation due to the beating heart and respiration. In this paper, we model the CS catheter as a set of electrodes. Novelly designed hypotheses generated by a number of learning-based detectors are fused. Robust hypothesis matching through a Bayesian framework is then used to select the best hypothesis for each frame. As a result, our tracking method achieves very high robustness against challenging scenarios such as low SNR, occlusion, foreshortening, non-rigid deformation, as well as the catheter moving in and out of ROI. Quantitative evaluation has been conducted on a database of 13221 frames from 1073 sequences. Our approach obtains 0.50mm median error and 0.76mm mean error. 97.8% of evaluated data have errors less than 2.00mm. The speed of our tracking algorithm reaches 5 frames-per-second on most data sets. Our approach is not limited to the catheters inside the CS but can be extended to track other types of catheters, such as ablation catheters or circumferential mapping catheters.Keywords
This publication has 10 references indexed in Scilit:
- Learning-based 3D myocardial motion flowestimation using high frame rate volumetric ultrasound data2010
- Four-Chamber Heart Modeling and Automatic Segmentation for 3-D Cardiac CT Volumes Using Marginal Space Learning and Steerable FeaturesIEEE Transactions on Medical Imaging, 2008
- Hierarchical Learning of Curves Application to Guidewire Localization in Fluoroscopy2007
- HRS/EHRA/ECAS Expert Consensus Statement on Catheter and Surgical Ablation of Atrial Fibrillation: Recommendations for Personnel, Policy, Procedures and Follow-UpHeart Rhythm, 2007
- Catheter motion during atrial ablation due to the beating heart and respiration: Impact on accuracy and spatial referencing in three-dimensional mappingHeart Rhythm, 2007
- Ensemble TrackingIEEE Transactions on Pattern Analysis and Machine Intelligence, 2007
- On-line Boosting and Vision2006
- Robust Real-Time Face DetectionInternational Journal of Computer Vision, 2004
- Real-time tracking of non-rigid objects using mean shift2002
- An efficient method for finding the minimum of a function of several variables without calculating derivativesThe Computer Journal, 1964