Radio frequency-driven proton source with a back-streaming electron dump
Authors:
- 1 February 2010
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 81 (2) , 02B312
Abstract
This article describes an rf ion source with a back-streaming electron dump. A quartz tube, brazed to a metal plug at one end, is fused in the center of a flat quartz plate. rf power (at 13.6 MHz) is coupled to generate hydrogen plasma using a planar external antenna bonded to the window. Bonding the water-cooled rf antenna to the quartz window significantly lowers its temperature. The water-cooled metal plug serves as the back-streaming electron dump. At 1800 W, the current density of extracted hydrogen ions reaches approximately .
Keywords
This publication has 4 references indexed in Scilit:
- Experiments with planar inductive ion source meant for creation of H+ beamsReview of Scientific Instruments, 2007
- Multicusp ion source with external rf antenna for production of protonsReview of Scientific Instruments, 2004
- Mini rf-driven ion sources for focused ion beam systemsReview of Scientific Instruments, 2003
- Improvement in brightness of multicusp-plasma ion sourceJournal of Vacuum Science & Technology B: Microelectronics and Nanometer Structures Processing, Measurement, and Phenomena, 2002