Automatic 3D registration of dynamic stress and rest 82Rb and flurpiridaz F 18 myocardial perfusion PET data for patient motion detection and correction
- 31 October 2011
- journal article
- nuclear medicine-physics
- Published by Wiley in Medical Physics
- Vol. 38 (11) , 6313-6326
Abstract
Purpose:: The authors aimed to develop an image‐based registration scheme to detect and correct patient motion in stress and rest cardiac positron emission tomography (PET)/CT images. The patient motion correction was of primary interest and the effects of patient motion with the use of flurpiridaz F 18 and82Rb were demonstrated.Methods:: The authors evaluated stress/rest PET myocardial perfusion imaging datasets in 30 patients (60 datasets in total, 21 male and 9 female) using a new perfusion agent (flurpiridaz F 18) (n = 16) and 82Rb (n = 14), acquired on a Siemens Biograph‐64 scanner in list mode. Stress and rest images were reconstructed into 4 (82Rb) or 10 (flurpiridaz F 18) dynamic frames (60 s each) using standard reconstruction (2D attenuation weighted ordered subsets expectation maximization). Patient motion correction was achieved by an image‐based registration scheme optimizing a cost function using modified normalized cross‐correlation that combined global and local features. For comparison, visual scoring of motion was performed on the scale of 0 to 2 (no motion, moderate motion, and large motion) by two experienced observers.Results:: The proposed registration technique had a 93% success rate in removing left ventricular motion, as visually assessed. The maximum detected motion extent for stress and rest were 5.2 mm and 4.9 mm for flurpiridaz F 18 perfusion and 3.0 mm and 4.3 mm for82Rb perfusion studies, respectively. Motion extent (maximum frame‐to‐frame displacement) obtained for stress and rest were (2.2 ± 1.1, 1.4 ± 0.7, 1.9 ± 1.3) mm and (2.0 ± 1.1, 1.2 ±0 .9, 1.9 ± 0.9) mm for flurpiridaz F 18 perfusion studies and (1.9 ± 0.7, 0.7 ± 0.6, 1.3 ± 0.6) mm and (2.0 ± 0.9, 0.6 ± 0.4, 1.2 ± 1.2) mm for 82Rb perfusion studies, respectively. A visually detectable patient motion threshold was established to be ≥2.2 mm, corresponding to visual user scores of 1 and 2. After motion correction, the average increases in contrast‐to‐noise ratio (CNR) from all frames for larger than the motion threshold were 16.2% in stress flurpiridaz F 18 and 12.2% in rest flurpiridaz F 18 studies. The average increases in CNR were 4.6% in stress 82Rb studies and 4.3% in rest 82Rb studies.Conclusions:: Fully automatic motion correction of dynamic PET frames can be performed accurately, potentially allowing improved image quantification of cardiac PET data.Keywords
This publication has 47 references indexed in Scilit:
- Simplified Quantification of Myocardial Flow Reserve with flurpiridaz F 18: Validation with Microspheres in a Pig ModelJournal of Nuclear Medicine, 2011
- Bias in iterative reconstruction of low-statistics PET data: benefits of a resolution modelPhysics in Medicine & Biology, 2011
- Single-Phase CT Aligned to Gated PET for Respiratory Motion Correction in Cardiac PET/CTJournal of Nuclear Medicine, 2010
- Enhanced definition PET for cardiac imagingJournal of Nuclear Cardiology, 2010
- Evaluation of the Novel Myocardial Perfusion Positron-Emission Tomography Tracer 18 F-BMS-747158-02Circulation, 2009
- Initial Characterization of an 18F-Labeled Myocardial Perfusion TracerJournal of Nuclear Medicine, 2008
- Clinical Myocardial Perfusion PET/CTJournal of Nuclear Medicine, 2007
- Mutual-information-based registration of medical images: a surveyIEEE Transactions on Medical Imaging, 2003
- Fast reconstruction of 3D PET data with accurate statistical modelingIEEE Transactions on Nuclear Science, 1998
- A Coefficient of Agreement for Nominal ScalesEducational and Psychological Measurement, 1960