Material and Biofilm Load of K Wires in Toe Surgery: Titanium versus Stainless Steel
Authors:
- 1 July 2013
- journal article
- research article
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 471 (7) , 2312-2317
Abstract
Recurrence rates for toe deformity correction are high and primarily are attributable to scar contractures. These contractures may result from subclinical infection. We hypothesized that (1) recurrence of toe deformities and residual pain are related to low-grade infections from biofilm formation on percutaneous K wires, (2) biofilm formation is lower on titanium (Ti) K wires compared with stainless steel (SS) K wires, and (3) clinical outcome is superior with the use of Ti K wires compared with SS K wires. In this prospective nonrandomized, comparative study, we investigated 135 lesser toe deformities (61 patients; 49 women; mean ± SD age, 60 ± 15 years) temporarily fixed with K wires between August 2010 and March 2011 (81 SS, 54 Ti). K wires were removed after 6 weeks. The presence of biofilm-related infections was analyzed by sonication. High bacterial loads (> 500 colony-forming units [CFU]/mL) were detected on all six toes requiring revision before 6 months. Increased bacterial load was associated with pain and swelling but not recurrence of the deformity. More SS K wires had greater than 100 CFU/mL bacteria than Ti K wires. For K wires with a bacterial count greater than 100 CFU/mL, toes with Ti K wires had a lower recurrence rate, less pain, and less swelling than toes with SS K wires. Ti K wires showed superior clinical outcomes to SS K wires. This appears to be attributable to reduced infection rates. Although additional study is needed, we currently recommend the use of Ti K wires for the transfixation of toe deformities. Level II, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.Keywords
This publication has 21 references indexed in Scilit:
- Biofilm formation on bone grafts and bone graft substitutes: Comparison of different materials by a standard in vitro test and microcalorimetryActa Biomaterialia, 2010
- The Most Current Algorithms for the Treatment and Prevention of Hypertrophic Scars and KeloidsPlastic and Reconstructive Surgery, 2010
- Sonication is superior to scraping for retrieval of bacteria in biofilm on titanium and steel surfaces in vitroActa Orthopaedica, 2009
- Titanium Alloy Pins Versus Stainless Steel Pins in External Fixation at the Wrist: A Randomized Prospective StudyJournal of Trauma: Injury, Infection & Critical Care, 2008
- Sonication of Removed Hip and Knee Prostheses for Diagnosis of InfectionThe New England journal of medicine, 2007
- Retrospective Study of the Association Between Hypertrophic Burn Scarring and Bacterial ColonizationJournal of Burn Care & Research, 2007
- Prosthetic-joint-associated infectionsBest Practice & Research Clinical Rheumatology, 2006
- Prosthetic-Joint InfectionsThe New England journal of medicine, 2004
- Adhesion of Staphylococcus to orthopaedic metals, an in vivo studyJournal of Orthopaedic Research, 2004
- Infection after intramedullary nailing: An experimental investigation on rabbitsInjury, 1996