Bedside Hemoglobinometry in Hemodialysis Patients: Lessons from Point-of-Care Testing
Authors:
- Rajiv AgarwalRARajiv Agarwal
- Tracy HeinzTHTracy Heinz
- 1 May 2001
- journal article
- research article
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 47 (3) , 240-243
Abstract
The HemoCue B-hemoglobin test system (HemoCue, Inc., Mission Viejo, CA) is a photometric method for rapid bedside determination of hemoglobin (Hb). We compared the performance of HemoCue measured Hb against Coulter STK-S (CSTK) measured Hb in chronic hemodialysis (HD) patients in two different settings. In the first setting, Hemocue analysis was performed by multiple HD technicians (n = 132). In the second setting, a nurse trained in proper specimen handling performed the HemoCue analysis (n = 74). Simultaneous measurement of Hb by the CSTK method was performed. First setting: Hb was 11.1 ± 1.66 (SD) g/dl by CSTK and 11.7 ± 2.29 g/dl by HemoCue. The HemoCue method consistently overestimated Hb by an average (SD) of 0.63 (1.267) g/dl (95% CI = 0.42 to 0.85). Hb was overestimated in 25.7% and underestimated in 2.3% of the patients by 1 g/dl or more. Thus, the HemoCue system was accurate within 1 g/dl only 72% of the time. Second setting: HemoCue overestimated Hb by an average (SD) of 0.29 (0.52) g/dl (95% CI, 0.17 to 0.41). Only 4% of all patients had errors in estimation of 1 g/dl or more. Thus, HemoCue was accurate in 96% of the patients within 1 g/dl. After reviewing the two protocols, the primary difference in the two studies was the technique used to obtain the specimens. When performed properly, Hb testing using the HemoCue testing system had a high level of agreement with CSTK. Appropriate training in specimen handling, as well as test performance, will increase accuracy and reliability of bedside hemoglobinometry.Keywords
This publication has 8 references indexed in Scilit:
- Linoleyl hydroperoxide transcriptionally upregulates heme oxygenase-1 gene expression in human renal epithelial and aortic endothelial cells.Journal of the American Society of Nephrology, 1998
- Measurement of haemoglobin using single drops of skin puncture blood: is precision acceptable?Journal of Clinical Pathology, 1998
- Application of the Department of Health and Human Services proposed waived status requirements for in vitro diagnostic testing devices: case studyClinical Chemistry, 1997
- HemoCue®, an Accurate Bedside Method of Hemoglobin Measurement?Journal of Clinical Monitoring and Computing, 1997
- A Clinical Evaluation of the Hemocue Haemoglobinometer using Capillary, Venous and Arterial SamplesAnaesthesia and Intensive Care, 1992
- Evaluation of portable haemoglobinometer in general practice.BMJ, 1987
- Evaluation of "HemoCue," a new device for determining hemoglobin.Clinical Chemistry, 1986
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986