A continuous flow mass spectrometry technique of argon isotope measurement for K/Ar geochronology
- 14 July 2009
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 23 (16) , 2403-2410
Abstract
A new method for the measurement of argon isotope composition in a continuous flow of helium for potassium/argon geochronology is described. Extraction of argon from geological samples in multiple‐sample holders was carried out in a chamber by heating with a continuous Nd‐YAG laser. The extracted and pre‐concentrated argon is passed through a chromatographic capillary column in a flow of helium. Argon is separated from possible contaminants in the column and is injected through an open split into the ion source of an isotope ratio mass spectrometer. Measurement of the 36 Ar, 38 Ar and 40 Ar isotopes was carried out in dynamic mode, using a triple‐collector ion detector. These experiments have shown that continuous flow mass spectrometry can be used for the analysis of radiogenic argon in picogram quantities with an accuracy that is satisfactory for the solution of many geochronological problems. The method of argon isotope measurement in a continuous flow of helium is an alternative to the measurement of argon isotopes in the static mode. The sensitivity and accuracy of argon measurement by this method are comparable with those provided by the classical static method. The measurement of argon isotopes in a continuous flow of helium is simpler and more reliable than measurement in the static mode. Copyright © 2009 John Wiley & Sons, Ltd.Keywords
This publication has 20 references indexed in Scilit:
- Coupled N, C and S stable isotope measurements using a dual‐column gas chromatography systemRapid Communications in Mass Spectrometry, 2007
- δ 34 S measurements on organic materials by continuous flow isotope ratio mass spectrometryRapid Communications in Mass Spectrometry, 2005
- Continuous flow 2 H/ 1 H and 18 O/ 16 O analysis of water samples with dual inlet precisionRapid Communications in Mass Spectrometry, 2004
- Mass Spectrometer Hardware for Analyzing Stable Isotope Ratios2004
- A rapid method for determination of hydrogen and oxygen isotope ratios from water and hydrous mineralsChemical Geology, 2001
- ConFlo III - an interface for high precision δ13C and δ15N analysis with an extended dynamic rangeRapid Communications in Mass Spectrometry, 1999
- High Precision Isotope Ratio Monitoring Techniques in Mass SpectrometryJournal of Mass Spectrometry, 1996
- Isotope-ratio-monitoring gas chromatography-mass spectrometryAnalytical Chemistry, 1978
- A refined ultrahigh-vacuum furnace for rare gas analysisJournal of Physics E: Scientific Instruments, 1978
- High Sensitivity Mass Spectrometer for Noble Gas AnalysisReview of Scientific Instruments, 1956