Abstract
We performed a conventional, fluoroscopy-guided ablation of ongoing typical atrial flutter at the preferred isthmus site. Despite 150 min of procedure time (using a 4 mm, irrigated tip catheter with settings of 43°C, 40 W and irrigation rate of 40 mL/min), block on the isthmus was not achieved. No atrial signals were left on the line of ablation. Through the left femoral vein, we introduced an intracardiac echocardiography probe and found a highly prominent Eustachian valve (EV; see Figure 1 and Supplementary material online, Movie S1 ) with high-voltage atrial potential at the edge of the EV. After ablation at this point, the flutter was terminated and bidirectional block was confirmed.