In situ observation of surface oxide layers on medical grade Ni‐Ti alloy during straining
Authors:
- 2 April 2008
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research Part A
- Vol. 88A (4) , 1000-1009
Abstract
Medical grade Ni‐Ti alloys with shape memory or pseudo‐elastic behavior exhibit good biocompatibility because of an electrochemically passive oxide layer on the surface. In this work, the mechanical stability of surface oxide layers is investigated during reversible pseudo‐elastic deformation of commonly applied medical grade Ni‐Ti wires. Surface oxide layers with varying thickness were generated by varying annealing times under air atmosphere. The thicknesses of the surface oxide layers were determined by means of Rutherford backscattering spectrometry. In situ scanning electron microscopy investigations reveal a damage mechanism, which is assumed to have a significant influence on the biocompatibility of the material. The conditions that lead to the appearance of cracks in the surface oxide layer or to the flaking of surface oxide layer particles are identified. The influence of the thickness of the surface oxide layer on the damage mode is characterized. The possible impact of the damaged surface oxide layer on the material's biocompatibility and the potentials to reduce or avoid the damage are discussed. © 2008 Wiley Periodicals, Inc. J Biomed Mater Res, 2009Keywords
This publication has 30 references indexed in Scilit:
- The release of nickel from orthodontic NiTi wires is increased by dynamic mechanical loading but not constrained by surface nitridationJournal of Biomedical Materials Research Part A, 2007
- New oxidation treatment of NiTi shape memory alloys to obtain Ni‐free surfaces and to improve biocompatibilityJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2005
- On the origin of Lüders-like deformation of NiTi shape memory alloysJournal of the Mechanics and Physics of Solids, 2005
- Biocompatibility and hemocompatibility of surface‐modified NiTi alloysJournal of Biomedical Materials Research Part A, 2003
- Effect of chemical etching and aging in boiling water on the corrosion resistance of nitinol wires with black oxide resulting from manufacturing processJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2003
- First closure of a large patent ductus arteriosus in an infant with an angulated nitinol plugCatheterization and Cardiovascular Interventions, 2002
- Effect of surface treatment of NiTi alloy on its corrosion behavior in Hanks’ solutionJournal of Biomedical Materials Research, 2002
- Initiation and propagation of localized deformation in elasto-plastic strips under uniaxial tensionInternational Journal of Plasticity, 1997
- Memory metal stents for palliation of malignant obstruction of the oesophagus and cardia.Gut, 1995
- Shape memory alloys in minimally invasive therapyMinimally Invasive Therapy, 1995