Simultaneous bright‐ and black‐blood whole‐heart MRI for noncontrast enhanced coronary lumen and thrombus visualization
Open Access
- 19 July 2017
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 79 (3) , 1460-1472
Abstract
Purpose: To develop a 3D whole‐heart Bright‐blood and black‐blOOd phase SensiTive (BOOST) inversion recovery sequence for simultaneous noncontrast enhanced coronary lumen and thrombus/hemorrhage visualization.Methods: The proposed sequence alternates the acquisition of two bright‐blood datasets preceded by different preparatory pulses to obtain variations in blood/myocardium contrast, which then are combined in a phase‐sensitive inversion recovery (PSIR)‐like reconstruction to obtain a third, coregistered, black‐blood dataset. The bright‐blood datasets are used for both visualization of the coronary lumen and motion estimation, whereas the complementary black‐blood dataset potentially allows for thrombus/hemorrhage visualization. Furthermore, integration with 2D image‐based navigation enables 100% scan efficiency and predictable scan times. The proposed sequence was compared to conventional coronary MR angiography (CMRA) and PSIR sequences in a standardized phantom and in healthy subjects. Feasibility for thrombus depiction was tested ex vivo.Results: With BOOST, the coronary lumen is visualized with significantly higher (P < 0.05) contrast‐to‐noise ratio and vessel sharpness when compared to conventional CMRA. Furthermore, BOOST showed effective blood signal suppression as well as feasibility for thrombus visualization ex vivo.Conclusion: A new PSIR sequence for noncontrast enhanced simultaneous coronary lumen and thrombus/hemorrhage detection was developed. The sequence provided improved coronary lumen depiction and showed potential for thrombus visualization. Magn Reson Med 79:1460–1472, 2018. © 2017 International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.Keywords
Funding Information
- Engineering and Physical Sciences Research Council (EP/N009258/1, EP/P001009/1, EP/P007619/1)
- Medical Research Council (MR/L009676/1)
- FONDECYT (N° 1161051)
This publication has 53 references indexed in Scilit:
- Nonrigid autofocus motion correction for coronary MR angiography with a 3D cones trajectoryMagnetic Resonance in Medicine, 2013
- Detection of Intracoronary Thrombus by Magnetic Resonance Imaging in Patients With Acute Myocardial InfarctionCirculation, 2011
- 3D flow‐independent peripheral vessel wall imaging using T2‐prepared phase‐sensitive inversion‐recovery steady‐state free precessionJournal of Magnetic Resonance Imaging, 2010
- Assessment of left ventricular myocardial scar in infiltrative and non-ischemic cardiac diseases by free breathing three dimensional phase sensitive inversion recovery (PSIR) TurboFLASHThe International Journal of Cardiovascular Imaging, 2010
- T2 quantification for improved detection of myocardial edemaJournal of Cardiovascular Magnetic Resonance, 2009
- Selective coronary artery plaque visualization and differentiation by contrast-enhanced inversion prepared MRIEuropean Heart Journal, 2006
- Modified Look‐Locker inversion recovery (MOLLI) for high‐resolution T1 mapping of the heartMagnetic Resonance in Medicine, 2004
- Respiratory motion in coronary magnetic resonance angiography: A comparison of different motion modelsJournal of Magnetic Resonance Imaging, 2002
- Coronary Angiography with Magnetization‐Prepared T2 ContrastMagnetic Resonance in Medicine, 1995
- Compensatory Enlargement of Human Atherosclerotic Coronary ArteriesThe New England journal of medicine, 1987