Is submucosal bladder pressure monitoring feasible?
Open Access
- 29 January 2018
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 233 (1) , 100-113
Abstract
There has been recent interest in placing pressure-sensing elements beneath the bladder mucosa to facilitate chronic bladder pressure monitoring. Wired submucosal sensors with the wires passed through detrusor have been demonstrated in vivo, with limited chronic retention, potentially due to the cable tethering the detrusor. Published studies of submucosal implants have shown that high correlation coefficients between submucosal and lumen pressures can be obtained in caprine, feline, and canine models. We have developed a wireless pressure monitor and surgical technique for wireless submucosal implantation and present our initial chronic implantation study here. Pressure monitors were implanted (n = 6) in female calf models (n = 5). Five devices were implanted cystoscopically with a 25-French rigid cystoscope. One device was implanted suprapubically to test device retention with an intact mucosa. Wireless recordings during anesthetized cystometry simultaneous with catheter-based reference vesical pressure measurements during filling and manual bladder compressions were recorded. Individual analysis of normalised data during bladder compressions (n = 12) indicated high correlation (r = 0.85–0.94) between submucosal and reference vesical pressure. The healing response was robust over 4 weeks; however, mucosal erosion occurred 2–4 weeks after implantation, leading to device migration into the bladder lumen and expulsion during urination. Wireless pressure monitors may be successfully placed in a suburothelial position. Submucosal pressures are correlated with vesical pressure, but may differ due to biomechanical forces pressing on an implanted sensor. Fully wireless devices implanted beneath the mucosa have risk of erosion through the mucosa, potentially caused by disruption of blood flow to the urothelium, or an as-yet unstudied mechanism of submucosal regrowth. Further investigation into device miniaturisation, anchoring methods, and understanding of submucosal pressure biomechanics may enable chronic submucosal pressure monitoring. However, the risk of erosion with submucosal implantation highlights the need for investigation of devices designed for chronic intravesical pressure monitoring.Keywords
Funding Information
- U.S. Department of Veterans Affairs (F7422-R)
- U.S. Department of Veterans Affairs (F9261-L)
This publication has 54 references indexed in Scilit:
- Ambulant monitoring of bladder oxygenation and hemodynamics using wireless near-infrared spectroscopyCanadian Urological Association Journal, 2013
- Development of Implantable Medical Devices: From an Engineering PerspectiveInternational Neurourology Journal, 2013
- Design and fabrication of implantable pressure sensing resistor sensor for human bladder monitoring systemMicrosystem Technologies, 2011
- The BladderPill: An in-body system logging bladder pressureSensors and Actuators A: Physical, 2010
- The additional value of ambulatory urodynamic measurements compared with conventional urodynamic measurementsBJU International, 2010
- Development of a fully implantable wireless pressure monitoring systemBiomedical Microdevices, 2008
- Risk factors for urinary stone formation in men with spinal cord injury: a 17‐year follow‐up studyBJU International, 2006
- Primer: questionnaires versus urodynamics in the evaluation of lower urinary tract dysfunction—one, both or none?Nature Reviews Endocrinology, 2005
- The Neurogenic Overactive Bladder. Classification Based on UrodynamicsBritish Journal of Urology, 1989
- Corrosion of Implant MaterialsAnnual Review of Materials Science, 1976