Harmonizing SUVs in multicentre trials when using different generation PET systems: prospective validation in non-small cell lung cancer patients

Authors:

Abstract
Purpose We prospectively evaluated whether a strategy using point spread function (PSF) reconstruction for both diagnostic and quantitative analysis in non-small cell lung cancer (NSCLC) patients meets the European Association of Nuclear Medicine (EANM) guidelines for harmonization of quantitative values. Methods The NEMA NU-2 phantom was used to determine the optimal filter to apply to PSF-reconstructed images in order to obtain recovery coefficients (RCs) fulfilling the EANM guidelines for tumour positron emission tomography (PET) imaging (PSFEANM). PET data of 52 consecutive NSCLC patients were reconstructed with unfiltered PSF reconstruction (PSFallpass), PSFEANM and with a conventional ordered subset expectation maximization (OSEM) algorithm known to meet EANM guidelines. To mimic a situation in which a patient would undergo pre- and post-therapy PET scans on different generation PET systems, standardized uptake values (SUVs) for OSEM reconstruction were compared to SUVs for PSFEANM and PSFallpass reconstruction. Results Overall, in 195 lesions, Bland-Altman analysis demonstrated that the mean ratio between PSFEANM and OSEM data was 1.03 [95 % confidence interval (CI) 0.94–1.12] and 1.02 (95 % CI 0.90–1.14) for SUVmax and SUVmean, respectively. No difference was noticed when analysing lesions based on their size and location or on patient body habitus and image noise. Ten patients (84 lesions) underwent two PET scans for response monitoring. Using the European Organization for Research and Treatment of Cancer (EORTC) criteria, there was an almost perfect agreement between OSEMPET1/OSEMPET2 (current standard) and OSEMPET1/PSFEANM-PET2 or PSFEANM-PET1/OSEMPET2 with kappa values of 0.95 (95 % CI 0.91–1.00) and 0.99 (95 % CI 0.96–1.00), respectively. The use of PSFallpass either for pre- or post-treatment (i.e. OSEMPET1/PSFallpass-PET2 or PSFallpass-PET1/OSEMPET2) showed considerably less agreement with kappa values of 0.75 (95 % CI 0.67–0.83) and 0.86 (95 % CI 0.78–0.94), respectively. Conclusion Protocol-optimized images and compliance with EANM guidelines allowed for a reliable pre- and post-therapy evaluation when using different generation PET systems. These data obtained in NSCLC patients could be extrapolated to other solid tumours.

This publication has 40 references indexed in Scilit: