Biomechanics of a Fixed–Center of Rotation Cervical Intervertebral Disc Prosthesis
Open Access
- 1 December 2012
- journal article
- Published by International Journal of Spine Surgery in International Journal of Spine Surgery
- Vol. 6 (1) , 34-42
Abstract
Past in vitro experiments studying artificial discs have focused on range of motion. It is also important to understand how artificial discs affect other biomechanical parameters, especially alterations to kinematics. The purpose of this in vitro investigation was to quantify how disc replacement with a ball-and-socket disc arthroplasty device (ProDisc-C; Synthes, West Chester, Pennsylvania) alters biomechanics of the spine relative to the normal condition (positive control) and simulated fusion (negative control). Specimens were tested in multiple planes by use of pure moments under load control and again in displacement control during flexion-extension with a constant 70-N compressive follower load. Optical markers measured 3-dimensional vertebral motion, and a strain gauge array measured C4-5 facet loads. Range of motion and lax zone after disc replacement were not significantly different from normal values except during lateral bending, whereas plating significantly reduced motion in all loading modes (P < .002). Plating but not disc replacement shifted the location of the axis of rotation anteriorly relative to the intact condition (P < 0.01). Coupled axial rotation per degree of lateral bending was 25% ± 48% greater than normal after artificial disc replacement (P = .05) but 37% ± 38% less than normal after plating (P = .002). Coupled lateral bending per degree of axial rotation was 37% ± 21% less than normal after disc replacement (P < .001) and 41% ± 36% less than normal after plating (P = .001). Facet loads did not change significantly relative to normal after anterior plating or arthroplasty, except that facet loads were decreased during flexion in both conditions (P < .03). In all parameters studied, deviations from normal biomechanics were less substantial after artificial disc placement than after anterior plating.Keywords
This publication has 16 references indexed in Scilit:
- Biomechanics of a Novel Minimally Invasive Lumbar Interspinous SpacerOperative Neurosurgery, 2010
- History of cervical disc arthroplastyNeurosurgical Focus, 2009
- The use of surface strain data and a neural networks solution method to determine lumbar facet joint loads during in vitro spine testingJournal of Biomechanics, 2008
- Range of motion change after cervical arthroplasty with ProDisc-C and Prestige artificial discs compared with anterior cervical discectomy and fusionJournal of Neurosurgery: Spine, 2007
- Changes in adjacent-level disc pressure and facet joint force after cervical arthroplasty compared with cervical discectomy and fusionJournal of Neurosurgery: Spine, 2007
- The Spinal Lax Zone and Neutral ZoneJournal of Spinal Disorders, 1998
- Construction of Local Vertebral Coordinate Systems Using a Digitizing ProbeSpine, 1997
- An Apparatus for Applying Pure Nonconstraining Moments to Spine Segments In VitroSpine, 1995
- Biomechanical Evaluation of Spinal Fixation DevicesSpine, 1988
- Rigid body motion calculated from spatial co-ordinates of markersJournal of Biomechanics, 1980