Prevention of Infections Associated With Permanent Cardiac Antiarrhythmic Devices by Implementation of a Comprehensive Infection Control Program
Authors:
- Jacob GiladJGJacob Gilad
- Eytan HyamEHEytan Hyam
- Pnina SchlaefferPSPnina Schlaeffer
- Amos KatzAKAmos Katz
- 1 June 2004
- journal article
- Published by Cambridge University Press (CUP) in Infection Control & Hospital Epidemiology
- Vol. 25 (6) , 492-497
Abstract
Objective: To implement a comprehensive infection control (IC) program for prevention of cardiac device-associated infections (CDIs).Design: Prospective before-after trial with 2 years of follow-up.Setting: A tertiary-care, university-affiliated medical center.Patients: A consecutive sample of all adults undergoing cardiac device implantation between 1997 and 2002.Intervention: An IC program was implemented during late 2001 and included staff education, preoperative modification of patient risk factors, intraoperative control of strict aseptic technique, surgical scrubbing and attire, control of environmental risk factors, optimization of antibiotic prophylaxis, postoperative wound care, and active surveillance. The clinical endpoint was CDI rates.Results: Between 1997 and 2000, there were 7 CDIs among 725 procedures (mean annual CDI incidence, 1%). During the first 9 months of 2001, there were 7 CDIs among 167 procedures (4.2%; P = .007): CDIs increased from 7 among 576 to 3 among 124 following pacemaker implantation (P = .39) and from 0 among 149 to 4 among 43 following cardioverter-defibrillator implantation (P = .002). Of the 14 CDIs, 5 involved superficial wounds, 7 involved deep wounds, and 2 involved endocarditis. Following intervention, there were no cases of CDI among 316 procedures during 24 months of follow-up (4.2% reduction; P = .0005).Conclusions: We observed a high CDI rate associated with substantial morbidity. IC measures had an impact on CDI. Although the relative weight of each measure in the prevention of CDI remains unknown, our results suggest that implementation of a comprehensive IC program is feasible and efficacious in this setting.Keywords
This publication has 28 references indexed in Scilit:
- ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices: Summary ArticleCirculation, 2002
- Nosocomial cardiac infectionsJournal of Hospital Infection, 2002
- Global Consensus Conference: Final recommendationsAmerican Journal of Infection Control, 1999
- Guideline for Prevention of Surgical Site Infection, 1999American Journal of Infection Control, 1999
- Requirements for Infrastructure and Essential Activities of Infection Control and Epidemiology in Hospitals: A Consensus Panel ReportInfection Control & Hospital Epidemiology, 1998
- Analysis of risk factors for nosocomial infections—results from the first national prevalence survey in Germany (NIDEP study, part 1)Journal of Hospital Infection, 1997
- APIC guidelines for handwashing and hand antisepsis in health care settingsAmerican Journal of Infection Control, 1995
- Surveillance of wound infectionsJournal of Hospital Infection, 1995
- Complications of Permanent Transvenous Cardiac Pacing: A 14‐Year Review of all Transvenous Pacemakers Inserted at One Community HospitalPacing and Clinical Electrophysiology, 1987
- COAGULASE‐NEGATIVE STAPHYLOCOCCI AND CLOXACILLIN PROPHYLAXIS IN PACEMAKER SURGERYActa Pathologica Microbiologica Scandinavica Series B: Microbiology, 1983