[Principle and results of a new "non-contact-tonometer" are discussed (author's transl)].

Authors:
  • JDJ Draeger
    J Draeger
  • KJK Jessen
    K Jessen
  • GHG Haselmann
    G Haselmann

  • 1 July 1975
    • journal article
    • research article
    • Vol. 167  (1) , 27-34
    • PMID:1202268
Abstract
Comparing measurements with Non-Contact-Tonometer and handapplanation tonometer on 462 eyes with normal and elevated intraocular tension are statistically evaluated. The 2-s-limits for normal i.o. pressure at +/- 6.52 mm Hg (+/- 8.6 mbar), for elevated intraocular pressure +/- 13.2 mm Hg (+/- 17.6 mbar). This means less reliability than conventional applanation tonometers or even Schiötz-tonometers. The unability to read the tension when the corneal surface is rough means remarkable disadvantages to clinical use.