Chemiluminescence Determination of Sulfur Dioxide in Air Using Tris(1,10-Phenanthroline)Ruthenium−KIO4System

Authors:
  • ZHZhike He
    Zhike He
  • FWFengwu Wu
    Fengwu Wu
  • HMHui Meng
    Hui Meng
  • LYLiangjie Yuan
    Liangjie Yuan
  • QLQingyao Luo
    Qingyao Luo

Abstract
The emission produced by sulfite in its oxidation by periodate in acidic solution in the presence of Ru(phen)3 2+ is used to determine 1.0x10−7 to 1.5x10−4 mol/L sulfite. The limit of detection is 7x10−10 mol/L and the relative standard deviation is 2.3% for a 4x10−5 mol/L sulfite solution (n=9). The method was applied satisfactorily to the determination of sulfur dioxide in air by using triethanolamine (TEA) as absorbent material.