Low-dose preview for patient-specific, task-specific technique selection in cone-beam CT
- 25 June 2014
- journal article
- Published by Wiley in Medical Physics
- Vol. 41 (7) , 071915
Abstract
A method is presented for generating simulated low-dose cone-beam CT (CBCT) preview images from which patient- and task-specific minimum-dose protocols can be confidently selected prospectively in clinical scenarios involving repeat scans. In clinical scenarios involving a series of CBCT images, the low-dose preview (LDP) method operates upon the first scan to create a projection dataset that accurately simulates the effects of dose reduction in subsequent scans by injecting noise of proper magnitude and correlation, including both quantum and electronic readout noise as important components of image noise in flat-panel detector CBCT. Experiments were conducted to validate the LDP method in both a head phantom and a cadaveric torso by performing CBCT acquisitions spanning a wide dose range (head: 0.8-13.2 mGy, body: 0.8-12.4 mGy) with a prototype mobile C-arm system. After injecting correlated noise to simulate dose reduction, the projections were reconstructed using both conventional filtered backprojection (FBP) and an iterative, model-based image reconstruction method (MBIR). The LDP images were then compared to real CBCT images in terms of noise magnitude, noise-power spectrum (NPS), spatial resolution, contrast, and artifacts. For both FBP and MBIR, the LDP images exhibited accurate levels of spatial resolution and contrast that were unaffected by the correlated noise injection, as expected. Furthermore, the LDP image noise magnitude and NPS were in strong agreement with real CBCT images acquired at the corresponding, reduced dose level across the entire dose range considered. The noise magnitude agreed within 7% for both the head phantom and cadaveric torso, and the NPS showed a similar level of agreement up to the Nyquist frequency. Therefore, the LDP images were highly representative of real image quality across a broad range of dose and reconstruction methods. On the other hand, naïve injection ofuncorrelated noise resulted in strong underestimation of the true noise, which would lead to overly optimistic predictions of dose reduction. Correlated noise injection is essential to accurate simulation of CBCT image quality at reduced dose. With the proposed LDP method, the user can prospectively select patient-specific, minimum-dose protocols (viz., acquisition technique and reconstruction method) suitable to a particular imaging task and to the user's own observer preferences for CBCT scans following the first acquisition. The method could provide dose reduction in common clinical scenarios involving multiple CBCT scans, such as image-guided surgery and radiotherapy.Keywords
Funding Information
- NIH (F32EB017571)
- AAPM (2013 Research Seed Funding)
- Siemens Healthcare, XP Division
- National Institutes of Health (F32EB017571)
- American Association of Physicists in Medicine (2013 Research Seed Funding)
- Siemens Healthcare, XP Division
- National Institutes of Health (F32EB017571)
- American Association of Physicists in Medicine (2013 Research Seed Funding)
- Siemens Healthcare, XP Division
This publication has 49 references indexed in Scilit:
- Development and validation of a low dose simulator for computed tomographyEuropean Radiology, 2009
- Performance comparison between total variation (TV)-based compressed sensing and statistical iterative reconstruction algorithmsPhysics in Medicine & Biology, 2009
- Image reconstruction in circular cone-beam computed tomography by constrained, total-variation minimizationPhysics in Medicine & Biology, 2008
- Geometric calibration of a mobile C‐arm for intraoperative cone‐beam CTMedical Physics, 2008
- A Technique for Simulating the Effect of Dose Reduction on Image Quality in Digital Chest RadiographyJournal of Digital Imaging, 2008
- Noise simulation in cone beam CT imaging with parallel computingPhysics in Medicine & Biology, 2006
- Cone-beam reprojection using projection-matricesIEEE Transactions on Medical Imaging, 2003
- A framework for noise‐power spectrum analysis of multidimensional imagesMedical Physics, 2002
- Penalized weighted least-squares image reconstruction for dual energy X-ray transmission tomographyIEEE Transactions on Medical Imaging, 2000
- The noise power spectrum in computed X-ray tomographyPhysics in Medicine & Biology, 1978