The American Association of Physicists in Medicine (AAPM) formed Task Group 221 (TG‐221) to discuss a generalized commissioning process, quality management considerations, and clinical physics practice standards for ocular plaque brachytherapy. The purpose of this report is also, in part, to aid the clinician to implement recommendations of the AAPM TG‐129 report, which placed emphasis on dosimetric considerations for ocular brachytherapy applicators used in the Collaborative Ocular Melanoma Study (COMS). This report is intended to assist medical physicists in establishing a new ocular brachytherapy program and, for existing programs, in reviewing and updating clinical practices. The report scope includes photon‐ and beta‐emitting sources and source:applicator combinations. Dosimetric studies for photon and beta sources are reviewed to summarize the salient issues and provide references for additional study. The components of an ocular plaque brachytherapy quality management program are discussed, including radiation safety considerations, source calibration methodology, applicator commissioning, imaging quality assurance tests for treatment planning, treatment planning strategies, and treatment planning system commissioning. Finally, specific guidelines for commissioning an ocular plaque brachytherapy program, clinical physics practice standards in ocular plaque brachytherapy, and other areas reflecting the need for specialized treatment planning systems, measurement phantoms, and detectors (among other topics) to support the clinical practice of ocular brachytherapy are presented. Expected future advances and developments for ocular brachytherapy are discussed.Keywords
This publication has 165 references indexed in Scilit:
- Quantifying the dosimetric influences of radiation coverage and brachytherapy implant placement uncertainty on eye plaque size selectionBrachytherapy, 2013
- 106Ruthenium Plaque Therapy (RPT) for RetinoblastomaInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012
- BEDVH-A method for evaluating biologically effective dose volume histograms: Application to eye plaque brachytherapy implantsMedical Physics, 2012
- Keeping an eye on the ring: COMS plaque loading optimization for improved dose conformity and homogeneityJournal of Contemporary Brachytherapy, 2012
- Palladium-103 Ophthalmic Plaque Radiation Therapy for Choroidal Melanoma: 400 Treated PatientsOphthalmology, 2009
- TU-C-AUD-06: Calculation of the Dose Distribution Around a Commercially Available 125I Brachytherapy Source Via a Multi-Group Discrete Ordinates MethodMedical Physics, 2007
- Finger's "slotted" eye plaque for radiation therapy: treatment of juxtapapillary and circumpapillary intraocular tumoursBritish Journal of Ophthalmology, 2007
- Beta-Ray Brachytherapy of Retinoblastoma: Feasibility of a New Small-Sized Ruthenium-106 PlaqueOphthalmic Research, 2005
- The three-dimensional scintillation dosimetry method: test for a106Ru eye plaque applicatorPhysics in Medicine & Biology, 2005
- Ruthenium Brachytherapy for Uveal Melanoma, 1979–2003Survival and Functional Outcomes in the Swedish PopulationOphthalmology, 2005