Design of Open High Magnetic Field MRI Superconducting Magnet With Continuous Current and Genetic Algorithm Method
Authors:
- Qiuliang WangQWQiuliang Wang
- Guoxin XuGXGuoxin Xu
- Yinming DaiYDYinming Dai
- Baozhi ZhaoBZBaozhi Zhao
- Luguang YanLYLuguang Yan
- Keeman KimKKKeeman Kim
- 23 June 2009
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Applied Superconductivity
- Vol. 19 (3) , 2289-2292
Abstract
An optimization design method of short-length actively shielded and open structure superconducting MRI magnets is suggested in the paper. Firstly, the section of the solenoid coil is simplified as a current loop with zero section to solve a linear programming problem. The position coordinates in the radius and axial, and current for the loop can be calculated by the linear programming method. Then, the cross-section of the coil is optimized with a genetic algorithm to get appropriate section size. The method of linear programming, especially combining with genetic algorithm, reduces optimizing variables, which makes the design of a magnet feasible. Based on the method, a full open MRI superconducting magnet is designed with maximum radii of 0.8 m and 1.2 m. In the paper, the detailed optimization technologies are presented.Keywords
This publication has 5 references indexed in Scilit:
- Superconducting Systems for MRI—Present Solutions and New TrendsIEEE Transactions on Applied Superconductivity, 2005
- A hybrid, inverse approach to the design of magnetic resonance imaging magnetsMedical Physics, 2000
- Homogeneous magnet design using linear programmingIEEE Transactions on Magnetics, 2000
- A study on optimal coil configurations in a split-type superconducting MRI magnetIEEE Transactions on Applied Superconductivity, 1999
- A methodological analysis of different formulations for solving inverse electromagnetic problemsIEEE Transactions on Magnetics, 1990