Ex vivo testing of the Quart® arterial line filter
- 1 December 1999
- journal article
- research article
- Published by SAGE Publications in Perfusion
- Vol. 14 (6) , 481-487
Abstract
Arterial line filters are now routinely used in cardiac surgery in order to decrease the microemboli load to the patient. The Quart® filter (Jostra, Hirrlingen, Germany) with a new planar construction design, an easy de-airing system and an integrated bypass, was tested for air filtration capacity and resistance to blood path in a standardized setting with surviving animals. Three calves (mean body weight: 71 ± 3.4 kg) were connected to a standard cardiopulmonary bypass (CPB) circuit by jugular venous and carotid arterial cannulation with a mean flow rate of 3.5 l/min. The arterial line filter was challenged with upstream injections of boluses of air of 5, 10 and 15 ml, respectively. A Doppler ultrasound was positioned downstream on the arterial line to measure bubble count and size. The pressure drop through the filter was monitored at flow rates of between 1 and 6 l/min. At the end of the procedure the animals were weaned from the CPB and, thereafter, from the ventilator. After 7 days, the animals were sacrificed electively. This study shows that important quantities of air can be injected into the arterial line upstream of the filter with small volumes of small sized bubbles recorded downstream. With the 5 ml air bolus injection, mean values of 0.3 ± 0.6 bubbles of 30 and 40 μm were detected, whereas with the 20 ml bolus, 32.6 ± 8.7 bubbles of 10 μm, 3.7 ± 1.1 bubbles of 30 μm, 3.3 ± 0.6 bubbles of 40 μm and 0.7 ± 1.1 bubbles of 50 μm were recorded. The blood path resistance at different blood flow rates was well within the acceptable range with a pressure drop of 20 ± 0 and 26.6 ± 5.7 mmHg at flow rates of 4 and 5 l/min, respectively. With its planar concept, the Quart filter offers good air filtering capacity both in terms of bubble count and size after injection of large boluses of air, without any increase of resistance to the blood path. Moreover, it offers a venting function and an integrated bypass system.Keywords
This publication has 18 references indexed in Scilit:
- Impact of Embolization During Coronary Artery Bypass Grafting on Outcome and Length of StayThe Annals of Thoracic Surgery, 1997
- Microemboli during coronary artery bypass grafting: Genesis and effect on outcomeThe Journal of Thoracic and Cardiovascular Surgery, 1995
- The impact of microemboli during cardiopulmonary bypass on neuropsychological functioning.Stroke, 1994
- When do cerebral emboli appear during open heart operations? A transcranial Doppler studyThe Annals of Thoracic Surgery, 1991
- Brain microemboli during cardiac surgery or aortographyAnnals of Neurology, 1990
- Filtration in cardiopulmonary bypass: past, present and futurePerfusion, 1988
- Cerebral microembolism during cardiopulmonary bypassThe Journal of Thoracic and Cardiovascular Surgery, 1988
- Evaluation of a range of arterial line filters: Part IPerfusion, 1987
- The Detection of Microemboli in the Middle Cerebral Artery during Cardiopulmonary Bypass: A Transcranial Doppler Ultrasound Investigation Using Membrane and Bubble OxygenatorsThe Annals of Thoracic Surgery, 1987
- Stroke Following Coronary Artery Bypass Grafting: A Ten-Year StudyThe Annals of Thoracic Surgery, 1985