Retinal phototoxicity observed using high-resolution autofluorescence imaging

Authors:

Abstract
The American National Standards Institute (ANSI) provides guidelines for the safe use of lasers from which maximum permissible exposures of laser light are calculated. This standard is used to limit ocular light exposures and to assess the safety of retinal imaging experiments. Using in vivo adaptive optics scanning laser ophthalmoscopy equipped with fluorescence imaging capabilities, we have observed retinal changes in the macaque retina caused by visible light exposures that are below the ANSI safety standard. Specifically, we observed an immediate decrease in the autofluorescence intensity of the retinal pigment epithelial cells at the exposure location, followed by long-term full recovery or permanent retinal damage. The AF reduction was observed with exposure to 488 nm or 568 nm light. We have demonstrated the photochemical nature of the AF reduction. Experimental evidence suggests that multiple lipofuscin fluorophores are involved in the reduction of AF. Since the mechanism causing the AF reduction is currently unknown, it remains unclear whether these effects are detrimental to retinal health. Even if the AF reduction is not hazardous, our results encourage a re-evaluation of the standards for avoiding photochemical damage from exposures in the middle of the visible spectrum.

This publication has 16 references indexed in Scilit: