Retinal phototoxicity observed using high-resolution autofluorescence imaging
Authors:
- 1 March 2009
- conference paper
- Published by Laser Institute of America
- Vol. 2009 (1) , 61-65
Abstract
The American National Standards Institute (ANSI) provides guidelines for the safe use of lasers from which maximum permissible exposures of laser light are calculated. This standard is used to limit ocular light exposures and to assess the safety of retinal imaging experiments. Using in vivo adaptive optics scanning laser ophthalmoscopy equipped with fluorescence imaging capabilities, we have observed retinal changes in the macaque retina caused by visible light exposures that are below the ANSI safety standard. Specifically, we observed an immediate decrease in the autofluorescence intensity of the retinal pigment epithelial cells at the exposure location, followed by long-term full recovery or permanent retinal damage. The AF reduction was observed with exposure to 488 nm or 568 nm light. We have demonstrated the photochemical nature of the AF reduction. Experimental evidence suggests that multiple lipofuscin fluorophores are involved in the reduction of AF. Since the mechanism causing the AF reduction is currently unknown, it remains unclear whether these effects are detrimental to retinal health. Even if the AF reduction is not hazardous, our results encourage a re-evaluation of the standards for avoiding photochemical damage from exposures in the middle of the visible spectrum.Keywords
This publication has 16 references indexed in Scilit:
- In Vivo Autofluorescence Imaging of the Human and Macaque Retinal Pigment Epithelial Cell MosaicInvestigative Ophthalmology & Visual Science, 2009
- In vivo fluorescence imaging of primate retinal ganglion cells and retinal pigment epithelial cellsOptics Express, 2006
- RETINAL INJURY THRESHOLDS FOR BLUE WAVELENGTH LASERSHealth Physics, 2006
- RPE Damage Thresholds and Mechanisms for Laser Exposure in the Microsecond-to-Millisecond Time RegimenInvestigative Ophthalmology & Visual Science, 2005
- Phototoxicity to the Retina: Mechanisms of DamageInternational Journal of Toxicology, 2002
- Local Fundus Response to Blue (LED and Laser) and Infrared (LED and Laser) SourcesExperimental Eye Research, 2001
- REVISION OF GUIDELINES ON LIMITS OF EXPOSURE TO LASER RADIATION OF WAVELENGTHS BETWEEN 400 nm AND 1.4 μmHealth Physics, 2000
- Distribution of fundus autofluorescence with a scanning laser ophthalmoscope.British Journal of Ophthalmology, 1995
- Retina: Ultrastructural Alterations Produced by Extremely Low Levels of Coherent RadiationScience, 1972
- An Equilibrium Thermal Model for Retinal Injury from Optical SourcesApplied Optics, 1969