Central Corneal Thickness Measurements With Different Imaging Devices and Ultrasound Pachymetry
Authors:
- Keat-Ween KhawKKKeat-Ween Khaw
- Choung-Min NgCNChoung-Min Ng
- Visvaraja SubrayanVSVisvaraja Subrayan
- 1 June 2013
- journal article
- research article
- Published by Wolters Kluwer Health in Cornea
- Vol. 32 (6) , 766-771
Abstract
The aim of this study was to compare 4 methods of central corneal thickness (CCT) measurements in terms of their agreement, repeatability, and measurement time. CCT was measured in 184 eyes of 92 healthy subjects by the same examiner. The methods used were as follows: noncontact specular microscopy (SM; Topcon SP-3000P; Topcon Corporation, Tokyo, Japan), Pentacam rotating Scheimpflug photography system (Oculus Inc, Wetzlar, Germany), optical low-coherence reflectometry (OLCR; LenStar LS900; Haag-Streit AG, Koeniz, Switzerland), and ultrasound pachymetry (UP; PachPen; Accutome Inc, Malvern, PA). The duration for each examination was measured by an independent observer. The mean age (±SD) of the subjects was 54.3 (± 15.3) years. The mean CCTs (± SD) for SM, Pentacam, OLCR, and UP were 507.8 (± 30.2), 538.4 (± 31.7), 531.8 (± 31.4), and 528.3 (± 32.9) μm, respectively. The Bland-Altman plots showed closest agreement for OLCR-UP, followed by OLCR-Pentacam and Pentacam-UP. SM had the poorest agreement with the other methods. CCTs measured by SM were on average 20 to 30 μm thinner than those of the other methods. The coefficient of repeatability for SM, Pentacam, OLCR, and UP were 3.14%, 4.23%, 1.51%, and 3.46%, respectively. The mean measurement times (± SD) were 13.5 (± 5.7), 45.7 (± 12.3), 18.5 (± 7.1), and 5.6 (± 1.0) seconds, respectively. CCT measurements between OLCR-UP and OLCR-Pentacam are comparable and can be used interchangeably in clinical practice. However, SM underestimates CCT compared with the other methods, whereas Pentacam was found to be the least repeatable and took the longest time.Keywords
This publication has 23 references indexed in Scilit:
- Comparison of central corneal thickness using optical low-coherence reflectometry and spectral-domain optical coherence tomographyJournal of Cataract & Refractive Surgery, 2012
- Comparison of the precision of the Topcon SP-3000P specular microscope and an ultrasound pachymeterClinical Ophthalmology, 2011
- Central Corneal Thickness Measured With Three Optical Devices and Ultrasound PachometryEye & Contact Lens: Science & Clinical Practice, 2011
- Comparison of Central Corneal Thickness Measurement Using Ultrasonic Pachymetry, Rotating Scheimpflug Camera, and Scanning-Slit TopographyAmerican Journal of Ophthalmology, 2010
- Corneal Thickness Measurement in the Management of Primary Open-angle GlaucomaOphthalmology, 2007
- Comparison of Central Corneal Thickness Measured by Four Different PachymetersEye & Contact Lens: Science & Clinical Practice, 2007
- Corneal thickness measurements in normal and keratoconic eyes: Pentacam comprehensive eye scanner versus noncontact specular microscopy and ultrasound pachymetryJournal of Cataract & Refractive Surgery, 2006
- Corneal thickness measurements: Scanning-slit corneal topography and noncontact specular microscopy versus ultrasonic pachymetryJournal of Cataract & Refractive Surgery, 2003
- Comparison of central corneal thickness measurements by specular microscopy, ultrasound pachymetry, and ultrasound biomicroscopyJournal of Cataract & Refractive Surgery, 2003
- Corneal thickness measurements with contact and noncontact specular microscopic and ultrasonic pachymetryAmerican Journal of Ophthalmology, 2001