Evaluation of corneal topographic, tomographic and biomechanical indices for detecting clinical and subclinical keratoconus: a comprehensive three-device study

Authors:

Abstract
AIM: To evaluate the diagnostic ability of topographic and tomographic indices with Pentacam and Sirius as well as biomechanical parameters with Corvis ST for the detection of clinical and subclinical forms of keratoconus (KCN). METHODS: In this prospective diagnostic test study, 70 patients with clinical KCN, 79 patients with abnormal findings in topography and tomography maps with no evidence on clinical examination (subclinical KCN), and 68 normal control subjects were enrolled. The accuracy of topographic, tomographic, and biomechanical parameters was evaluated using the area under the receiver operating characteristic curve (AUC) and cross-validation analysis. The Delong method was used for comparing AUCs. RESULTS: In distinguishing KCN from normal, all parameters showed statistically significant differences between the two groups (P0.05). CONCLUSION: In clinical KCN, all topographic, tomographic, and biomechanical indices have acceptable outcomes in terms of sensitivity and specificity. However, in differentiating subclinical forms of KCN from normal corneas, curvature-based parameters (SIb and IS value) followed by integrated indices (PRFI and TBI) are the most powerful tools for early detection of KCN.