AAEM minimonograph #16: Instrumentation and measurement in electrodiagnostic medicine–part I
Authors:
- 1 August 1995
- journal article
- review article
- Published by Wiley in Muscle & Nerve
- Vol. 18 (8) , 799-811
Abstract
Technical and instrumentation factors play an important role in obtaining reliable information during electrodiagnostic studies. With contemporary electrodiagnostic equipment, neurophysiologic potentials are detected using a variety of electrodes and undergo differential amplification, filtering, conversion to digital form, and finally, analysis and display. Understanding the signal processing principles, limitations, and sources of errors that can occur during this multistep process can improve the technical quality of studies, minimize preventable errors, and improve clinical interpretation. Part I of this minimonograph reviews the basic principles of action potential generation and overviews electrodiagnostic instrumentation. The concept of waveform frequency content is related to the role of filters in suppressing noise while preserving waveform latency, amplitude, and morphology. The electrical characteristics of various surface and needle electrodes influence instrument design and the nature of the potentials recorded. This is especially important in understanding the differences in motor unit characteristics obtained from monopolar and concentric needle electrodes. © 1995 John Wiley & Sons, Inc.Keywords
This publication has 18 references indexed in Scilit:
- The influence of the reference electrode on CMAP configurationMuscle & Nerve, 1993
- AAEM minimonograph #39: Digital filtering: Basic concepts and application to evoked potentialsMuscle & Nerve, 1992
- AAEM minimonograph #36: Basic concepts of electricity and electronics in clinical electromyographyMuscle & Nerve, 1991
- Recording characteristics of monopolar emg electrodesMuscle & Nerve, 1991
- Electronic Filter Effects on Normal Motor and Sensory Nerve Conduction TestsAmerican Journal of Physical Medicine & Rehabilitation, 1990
- Properties of motor unit action potentials recorded with concentric and monopolar needle electrodes: ADEMG analysisMuscle & Nerve, 1988
- Practical Instrumentation and Common Sources of ErrorAmerican Journal of Physical Medicine & Rehabilitation, 1988
- Motor Unit AnalysisAmerican Journal of Physical Medicine & Rehabilitation, 1988
- Simulation of concentric needle EMG motor unit action potentialsMuscle & Nerve, 1988
- Action Potential Parameters in Normal Human Muscle and their Dependence on Physical Variables.Acta Physiologica Scandinavica, 1954