Large single centre experience with the Cera™ and CeraFlex™ occluders for closure of interatrial communications: usefulness of the flexible rotation feature
- 10 November 2016
- journal article
- research article
- Published by Springer Nature in Cardiovascular Intervention and Therapeutics
- Vol. 33 (1) , 70-76
Abstract
We present our experience with the Cera (CO) and the CeraFlex occluder (CFO) in transcatheter closure of interatrial communications (IAC). Between 2013 and 2016, 201 patients (75 males, 16 with patent foramen ovale), aged 27 ± 19 (5–75) years, underwent percutaneous closure of IAC using CO or CFO in our institution. After transoesophageal imaging, the procedure was aborted in 7 young paediatric (6–13 years old) patients (3 multiple holes, 3 too small septum, 1 leak with 38 mm occlusion balloon). The occluder was removed prior to release in 11 patients (5.7%), while occlusion was successful in 183 patients (94.3%) with 44 CO, 136 CFO, and 3 Cera multifenestrated occluders. There were no deaths, embolizations, or major complications. Small residual shunts were demonstrated in 8 patients immediately after implantation, 4 (8.5%) with CO and 4 (2.9%) with CFO, all disappearing after 3 months. Over 1.8 ± 1.7 year follow-up, all patients improved with 2 asymptomatic, transient pericardial effusions and 5 adults with transient supraventricular arrhythmias, treated medically for 6 months. IAC closure with CO and CFO proved safe with favourable success rates and few, nonserious complications. The CFO flexible rotation feature helped in conforming to various septal anatomies, minimising manoeuvres and possibly post-occlusion leaks.Keywords
This publication has 15 references indexed in Scilit:
- Ceraflex versus Amplatzer occluder for secundum atrial septal defect closureHerz, 2015
- Holy Grail Deformity of Distal Disc Hubless Atrial Septal Occluders in Pediatric HeartsInternational Heart Journal, 2015
- A multicentre, comparative study of Cera septal occluder versus AMPLATZER Septal Occluder in transcatheter closure of secundum atrial septal defectsEuroIntervention, 2014
- Late erosion of Amplatzer septal occluder device resulting in cardiac tamponadeInteractive CardioVascular and Thoracic Surgery, 2014
- Atrial Septal Defect ClosureCardiology Clinics, 2013
- Tulip malformation of the left atrial disc in the lifetech cera ASD device: A novel complication of percutaneous ASD closureCatheterization and Cardiovascular Interventions, 2011
- Morphologic Characteristics and Relating Factors to the Need of Technical Modification in Transcatheter Closure of Large Atrial Septal Defect (≥5 mm)Korean Circulation Journal, 2010
- Long-term outcome of transcatheter secundum-type atrial septal defect closure using Amplatzer septal occludersJACC, 2005
- Erosion of Amplatzer septal occluder device after closure of secundum atrial septal defects: Review of registry of complications and recommendations to minimize future riskCatheterization and Cardiovascular Interventions, 2004
- Early clinical experience with use of the ‘Amplatzer Septal Occluder’ device for atrial septal defectCardiology in the Young, 1998