An eNose-based method performing drift correction for online VOC detection under dry and humid conditions
- 19 August 2020
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Analytical Methods
- Vol. 12 (39) , 4724-4733
Abstract
Our recently demonstrated innovative concept of electronic nose (eNose) based on a combination of gas sensors is further tested and benchmarked in the present study. The system is a test bed for gas sensors of different principal technologies distributed within three compartments, which share a compact, very low volume, temperature-controlled sensing chamber. Here, the eNose-based analyser contains three sensing arrays of commercially available semiconducting metal oxide (MOX) gas sensors: one compartment contains 8 analog MOX sensors, while the other two compartments comprise 10 digital MOX sensors. The presented instrument is explicitly tested for the discrimination between mid-range (3–18 ppm) concentrations of different volatile organic compounds (VOCs) including acetaldehyde, acetone, ethanol, ethyl acetate, isoprene and n-pentane under dry and humid conditions, which are all considered relevant gases in future breath diagnostic applications. Since the experiments were performed in periods of time separated by around 20 days, they are affected by drift. For this reason, we explore the opportunity of drift mitigation using methods based on component removal computed by linear discriminant analysis, partial least squares discriminant analysis and direct orthogonalization, which lend themselves to future in-field applications of the developed device and sensing methodology.Keywords
This publication has 63 references indexed in Scilit:
- Semiconducting metal oxides as sensors for environmentally hazardous gasesSensors and Actuators B: Chemical, 2011
- Metal Oxide Semi-Conductor Gas Sensors in Environmental MonitoringSensors, 2010
- Drift compensation of gas sensor array data by common principal component analysisSensors and Actuators B: Chemical, 2010
- Classification of Tempranillo wines according to geographic origin: Combination of mass spectrometry based electronic nose and chemometricsAnalytica Chimica Acta, 2010
- Geographical origin of Sauvignon Blanc wines predicted by mass spectrometry and metal oxide based electronic noseAnalytica Chimica Acta, 2009
- Template‐Free Synthesis and Assembly of Single‐Crystalline Tungsten Oxide Nanowires and their Gas‐Sensing PropertiesAngewandte Chemie, 2005
- Diagnostic potential of breath analysis—focus on volatile organic compoundsClinica Chimica Acta, 2004
- Accuracy and precision of flowing afterglow mass spectrometry for the determination of the deuterium abundance in the headspace of aqueous liquids and exhaled breath waterRapid Communications in Mass Spectrometry, 2001
- Analysis of discrimination mechanisms in the mammalian olfactory system using a model noseNature, 1982
- Zum Einflu von Wasserstoff auf die elektrische Leitf higkeit an der Oberfl che von ZinkoxydkristallenThe European Physical Journal A, 1957