This is the first in vivo kinematic study to compare mobile-bearing with fixed-bearing prostheses in patients who had total knee arthroplasties. Femorotibial contact positions for 40 patients implanted with either a fixed-bearing or mobile-bearing prosthesis were analyzed using videofluoroscopy. Femorotibial contact paths were determined using a computer automated model-fitting technique. Nineteen of 20 patients in each group experienced posterior femoral rollback of their lateral condyles, with a mean of 3.6 and 3.7 mm for fixed-bearing and mobile-bearing prostheses respectively. Eighteen patients who had mobile-bearing prostheses and 17 patients with fixed-bearing knee prostheses experienced a normal pattern of axial rotation of 7.3° and 4.1° respectively. Eleven of 20 (55%) patients who had mobile-bearing prostheses implanted and eight of 20 (40%) patients who had fixed-bearing prostheses implanted did not experience femoral condylar lift-off. The remaining knees had condylar lift-off less than 2.4 mm for fixed-bearing prostheses and 1.7 mm for mobile-bearing prostheses, respectively. Patients who had mobile-bearing prostheses implanted experienced greater axial rotation and less condylar lift-off than patients who had fixed-bearing prostheses implanted. Both cruciate ligaments are sacrificed for the mobile and fixed-bearing total knee replacements. The results from the current study showed that, in both groups, the majority of patients experienced kinematics similar to those of a normal knee. However, the extent of lateral femoral condyle posterior rollback and the extent of axial rotation were less.Keywords
This publication has 26 references indexed in Scilit:
- In Vivo Fluoroscopic Analysis Of Fixed-Bearing Total Knee ReplacementsClinical Orthopaedics and Related Research, 2003
- Kinematic comparison of posterior cruciate sacrifice versus substitution in a mobile bearing total knee arthroplastyThe Journal of Arthroplasty, 2002
- Femoral condylar lift-off in vivo in total knee arthroplastyThe Journal of Bone and Joint Surgery. British volume, 2001
- In Vivo Anteroposterior Femorotibial Translation of Total Knee ArthroplastyClinical Orthopaedics and Related Research, 1998
- Range of motion after total knee arthroplasty The effect of implant design and weight-bearing conditionsThe Journal of Arthroplasty, 1998
- In Vivo Knee Kinematics Derived Using an Inverse Perspective TechniqueClinical Orthopaedics and Related Research, 1996
- Patient Outcomes Following Tricompartmental Total Knee ReplacementJAMA, 1994
- Functional Analysis of Pre and Post-Knee Surgery: Total Knee Arthroplasty and ACL ReconstructionJournal of Biomechanical Engineering, 1993
- Interaction between intrinsic knee mechanics and the knee extensor mechanismJournal of Orthopaedic Research, 1987
- Knee biomechanics and total knee replacementThe Journal of Arthroplasty, 1986