Degradation of polyurethane foams used in the mêcme breast implant
Author:
- 1 October 1993
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 27 (10) , 1341-1348
Abstract
The release of toluene diisocyanate derivatives from polyurethane foam covers that were removed from the Měme breast implant was studied. The polyurethane foam covers as retrieved from the implants were contaminated on average with 2,4‐ and 2,6‐toluene diamine (TDA), 2,4‐ and 2,6‐toluene diisocyanate (TDI), and toluene isocyanate amine (TIA) at the 1086 ppm level (combined TDA, TDI, and TIA). When cleaned foam was incubated in Ringer's solution at 37°C for periods ranging from 6–35 days, 4.6 ppm/day of TDA, TDI, and TIA (combined) was formed; the projected annual degradation rate of the foam under these conditions was estimated to be about 0.8%. When cleaned foam was incubated dry at 37°C, 1.5 ppm/day of TDA, TDI, and TIA (combined) was produced. Hence, the polyurethane foam covers used in the Měme breast implants are susceptible to both thermal and hydrolytic degradation under simulated physiological conditions. © 1993 John Wiley & Sons, Inc.Keywords
This publication has 17 references indexed in Scilit:
- Release of Chemicals from Polyurethane Foam in the Meme Breast ImplantBiomaterials, Artificial Cells and Immobilization Biotechnology, 1993
- An Assessment of 2,4 TDA Formation from Surgitek Polyurethane Foam under Simulated Physiological ConditionsJournal of Biomaterials Applications, 1991
- An automated in vitro dermal absorption procedure: I. Permeation of 14C-labelled N,N-diethyl-m-toluamide through human skin and effects of short-wave ultraviolet radiation on permeationToxicology in Vitro, 1990
- Toxic hydrolysis product from a biodegradable foam implantJournal of Biomedical Materials Research, 1989
- Application of polyurethane foam for sampling volatile mutagens from ambient airScience of The Total Environment, 1987
- Detection of urinary amine metabolites in toluene diisocyanate exposed ratsJournal of Chromatography A, 1985
- Direct determination of isocyanates and amines as degradation products in the industrial production of polyurethane-coated wireThe Analyst, 1984
- Trace analysis of amines and isocyanates using glass capillary gas chromatography and selective detectionJournal of Chromatography A, 1983
- High-performance liquid chromatographic determination of 2,4- and 2,6-toluenediamine in aqueous extractsJournal of Chromatography A, 1982
- Determination of 2,4- and 2,6-diaminotoluene in flexible urethane foamsAnalytical Chemistry, 1977