Adoption of Exhaust Air Dust Testing in SPF Rodent Facilities
- 1 March 2020
- journal article
- research article
- Published by American Association for Laboratory Animal Science in Journal of the American Association for Laboratory Animal Science
- Vol. 59 (2) , 156-162
Abstract
Reliable detection of unwanted microbial agents is essential for meaningful health monitoring in laboratory animal facilities. Most rodents at our institution are housed in IVC rack systems to minimize aerogenic transmission of infectious agents between cages. The most commonly used rodent health monitoring systems expose live sentinel rodents to soiled bedding collected from other rodent cages on IVC racks and subsequently test these soiled-bedding sentinels for evidence of infection with excluded agents. However, infectious agents might go undetected when using this health surveillance method, due to inefficient organism shedding or transmission failure. In 2016, our institution switched the health monitoring methodology for the majority of our SPF rodent colonies to real-time PCR testing of environmental samples collected from the exhaust plenums of IVC racks. Here we describe our rationale for this conversion, describe some interesting health monitoring cases that arose soon after the conversion, and discuss a potential problem with the conversion—residual nucleic acids. We compared cost and implementation effort associated with 2 sampling methods, sticky swabs and in-line collection media. We also compared the ability of these 2 sampling methods to detect 2 prevalent microbes in our facilities, Helicobacter and murine norovirus. Our institution-wide switch to health monitoring by real-time PCR assay of exhaust air dust samples thus far has provided a sensitive, simple, and reliable approach for maintenance of SPF conditions in laboratory rodents and has dramatically reduced the use of live sentinel animals.Keywords
This publication has 20 references indexed in Scilit:
- FELASA recommendations for the health monitoring of mouse, rat, hamster, guinea pig and rabbit colonies in breeding and experimental unitsLaboratory Animals, 2014
- Efficacy of direct detection of pathogens in naturally infected mice by using a high-density PCR array.2013
- PCR Testing of a Ventilated Caging System to Detect Murine Fur Mites2013
- Age-associated variability in susceptibility of Swiss Webster mice to MPV and other excluded murine pathogens.2012
- Soiled bedding sentinels for the detection of fur mites in mice.2011
- Lymphocytic choriomeningitis infection undetected by dirty-bedding sentinel monitoring and revealed after embryo transfer of an inbred strain derived from wild mice.2007
- Failure of a soiled bedding sentinel system to detect cilia-associated respiratory bacillus infection in rats.1995
- The efficacy of a dirty bedding sentinel system for detecting Sendai virus infection in mice: a comparison of clinical signs and seroconversion.1994
- Age-dependent Susceptibility of Murine T Lymphocytes to Lymphocytic Choriomeningitis VirusJournal of General Virology, 1983
- Laboratory-Associated Infections: Incidence, Fatalities, Causes, and PreventionAnnual Review of Microbiology, 1979