Nodule detection in digital chest radiography: effect of system noise
Authors:
- Markus HåkanssonMarkus Håkansson
- Magnus BåthMagnus Båth
- Sara BörjessonSBSara Börjesson
- Susanne KheddacheSKSusanne Kheddache
- Åse Allansdotter JohnssonÅJÅse Allansdotter Johnsson
- Lars Gunnar MånssonLMLars Gunnar Månsson
- 17 May 2005
- journal article
- clinical trial
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 114 (1-3) , 97-101
Abstract
Apart from the image content that is the reproduction of anatomy and possible lesions, an X-ray image also contains system noise due to the limited number of photons and other internal noise sources in the system (image plate artefacts, electronic noise, etc.). The aim of this study was to determine the extent to which the system noise influences the detection of subtle lung nodules in five different regions of the chest. This was done by conducting a receiver operating characteristic (ROC) study with five observers on two different sets of images; clinical chest X-ray images and images of a LucAl phantom at similar dose levels found in the different regions of the chest. In both image types, mathematically simulated nodules (with a full-width-at-fifth-maximum of 10 mm) were added to the images at varying contrast levels. As a measure of the influence of system noise on the detection of subtle lung nodules, the ratio between the contrast needed to obtain an area under the ROC curve of 0.80 in the system noise images to that needed in the clinical images was used. The contrast ratio between system noise images and clinical images ranged from ∼0.02 (in the hilar region) to 0.18 (in the lower mediastinal region). The maximum difference in contrast needed for the corresponding system noise images, collected at the lowest and the highest dose represented in the anatomical image, was a factor of 2. These results indicate that probably no region in a chest X-ray image is limited by the number of quanta to the detector for the detection of 10 mm lung nodules when a radiation dose corresponding to a system with speed class 200 (leading to a detector dose of ∼9 μGy behind the parenchyma) is used.Keywords
This publication has 10 references indexed in Scilit:
- A software tool for increased efficiency in observer performance studies in radiologyRadiation Protection Dosimetry, 2005
- Nodule detection in digital chest radiography: introduction to the RADIUS chest trialRadiation Protection Dosimetry, 2005
- Nodule detection in digital chest radiography: summary of the RADIUS chest trialRadiation Protection Dosimetry, 2005
- Nodule detection in digital chest radiography: effect of anatomical noiseRadiation Protection Dosimetry, 2005
- Nodule detection in digital chest radiography: effect of nodule locationRadiation Protection Dosimetry, 2005
- Nodule detection in digital chest radiography: part of image background acting as pure noiseRadiation Protection Dosimetry, 2005
- Detection of Subtle Lung Nodules: Relative Influence of Quantum and Anatomic Noise on Chest RadiographsRadiology, 1999
- Nodule detection in two-component noise: toward patient structure1997
- Nodule Detection With and Without a Chest ImageInvestigative Radiology, 1985
- Beam quality independent attenuation phantom for estimating patient exposure from x-ray automatic exposure controlled chest examinationsMedical Physics, 1984