In-vivo perfusion quantification by contrast ultrasound: Validation of the use of linearized video data vs. raw RF data
Authors:
- 1 November 2008
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- Vol. 97 (10510117) , 1690-1693
Abstract
Perfusion quantification software commonly analyzes log-compressed video images provided by ultrasound systems. Compared to analysis of raw RF signals, errors in perfusion parameter estimates may be introduced, even when proper linearization is applied to reverse the log-compression. In this work, the effects of log-compression, demodulation, scan conversion and 8-bit quantization have been investigated, in terms of the errors introduced on perfusion parameter estimates. An in-vivo contrast bolus sequence was processed using both raw RF data and linearized video data. The outcome is that linearizing log-compressed video sequences can provide estimates as accurate as processing RF data under two conditions: a) system gain should be such as to prevent both excessive truncation of pixel values to minimum level and saturation to maximum level; b) the dynamic range of log-compression should be 45 dB or higher.Keywords
This publication has 4 references indexed in Scilit:
- A new formalism for the quantification of tissue perfusion by the destruction-replenishment method in contrast ultrasound imagingIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2006
- A multivessel model describing replenishment kinetics of ultrasound contrast agent for quantification of tissue perfusionUltrasound in Medicine & Biology, 2003
- FEMMINA: a fast echographic multiparametric multi-imaging novel apparatus1999
- Quantification of Myocardial Blood Flow With Ultrasound-Induced Destruction of Microbubbles Administered as a Constant Venous InfusionCirculation, 1998