Methods to produce calibration mixtures for anesthetic gas monitors and how to perform volumetric calculations on anesthetic gases
Authors:
- 1 October 1992
- journal article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 8 (4) , 279-284
Abstract
A simple procedure for making calibration mixtures of oxygen and the anesthetic gases isoflurane, enflurane, and halothane is described. One to ten grams of the anesthetic substance is evaporated in a closed, 11,361-cc glass bottle filled with oxygen gas at atmospheric pressure. The carefully mixed gas is used to calibrate anesthetic gas monitors. By comparison of calculated and measured volumetric results it is shown that at atmospheric conditions the volumetric behavior of anesthetic gas mixtures can be described with reasonable accuracy using the ideal gas law. A procedure is described for calculating the deviation from ideal gas behavior in cases in which this is needed.Keywords
This publication has 4 references indexed in Scilit:
- Evaluation of a multigas anaesthetic monitor: the Datex CapnomacAnaesthesia, 1988
- An Evaluation of Three Volatile Anaesthetic Agent MonitorsAnaesthesia and Intensive Care, 1986
- Evaluation of the Datex ‘Normac’ anaesthetic agent monitorAnaesthesia, 1985
- A laboratory investigation of a multigas monitor for anaesthesia (EMMA)Anaesthesia, 1982