Experimental Investigation of the Band Broadening Arising from Short-Range Interchannel Heterogeneities in Chromatographic Beds under the Condition of Identical External Porosity
- 11 December 2008
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 81 (2) , 705-715
Abstract
The interesting possibilities offered by micropillar array columns to investigate the relation between bed heterogeneity and band broadening in chromatographic columns are illustrated and investigated. The perfect control over the microscopic bed structure offered by the photolithographic fabrication technique could be used to produce a heterogeneous bed displaying a short-range interchannel heterogeneity and having exactly the same particle size and external porosity as the perfectly ordered bed that was used as the reference system. According to this approach, any observed difference in band broadening could be directly owed to the effect of the bed heterogeneity. Knowing that the only difference between the investigated beds was their degree of heterogeneity, the obtained plate height measurements provided a unique test to distinguish among the abilities to recognize short-range interchannel heterogeneity of the different existing plate height models. Interpreting the data with the classical van Deemter and Knox models, it was found that the increased bed heterogeneity leads not only to an increase in the A-term constant but also to a clear increase in the C-term constant.Keywords
This publication has 44 references indexed in Scilit:
- Packing density of slurry‐packed capillaries at low aspect ratiosJournal of Separation Science, 2008
- Pore-Scale Dispersion in Electrokinetic Flow through a Random Sphere PackingAnalytical Chemistry, 2006
- Domain Size-Induced Heterogeneity as Performance Limitation of Small-Domain Monolithic Columns and Other LC Support TypesAnalytical Chemistry, 2006
- A first principles explanation for the experimentally observed increase in A-term band broadening in small domain silica monoliths and other chromatographic supportsJournal of Chromatography A, 2005
- Influence of the packing heterogeneity on the performance of liquid chromatography supportsJournal of Chromatography A, 2005
- Influence of the Pillar Shape on the Band Broadening and the Separation Impedance of Perfectly Ordered 2-D Porous Chromatographic MediaAnalytical Chemistry, 2004
- Advantages of Perfectly Ordered 2-D Porous Pillar Arrays over Packed Bed Columns for LC Separations: A Theoretical AnalysisAnalytical Chemistry, 2003
- Simulation of Packed-Bed Chromatography Utilizing High-Resolution Flow Fields: Comparison with ModelsAnalytical Chemistry, 2002
- Nuclear magnetic resonance studies of solvent flow through chromatographic columns: effect of packing density on flow patternsJournal of Chromatography A, 2000
- Evidence on the Nature of Eddy Diffusion in Gas Chromatography from Inert (Nonsorbing) Column Data.Analytical Chemistry, 1963