Comparison of the Hemocompatibility of an Axial and a Centrifugal Left Ventricular Assist Device in an In Vitro Test Circuit
Authors:
- Patrick BorchersPBPatrick Borchers
- Patrick WinnersbachPWPatrick Winnersbach
- Sandra KraemerSKSandra Kraemer
- Christian BeckersCBChristian Beckers
- Eva Miriam BuhlEBEva Miriam Buhl
- Steffen LeonhardtSteffen Leonhardt
- Rolf RossaintRRRolf Rossaint
- Marian WalterMWMarian Walter
- Thomas BreuerTBThomas Breuer
- Christian BleilevensChristian Bleilevens
Open Access
- 15 June 2022
- journal article
- research article
- Published by MDPI AG in Journal of Clinical Medicine
- Vol. 11 (12) , 3431
Abstract
Background: Hemocompatibility of left ventricular assist devices is essential for preventing adverse events. In this study, we compared the hemocompatibility of an axial-flow (Sputnik) to a centrifugal-flow (HeartMate 3) pump. Methods: Both pumps were integrated into identical in vitro test circuits, each filled with 75 mL heparinized human blood of the same donor. During each experiment (n = 7), the pumps were operated with equal flow for six hours. Blood sampling and analysis were performed on a regular schedule. The analytes were indicators of hemolysis, coagulation activation, platelet count and activation, as well as extracellular vesicles. Results: Sputnik induced higher hemolysis compared to the HeartMate 3 after 360 min. Furthermore, platelet activation was higher for Sputnik after 120 min onward. In the HeartMate 3 circuit, the platelet count was reduced within the first hour. Furthermore, Sputnik triggered a more pronounced increase in extracellular vesicles, a potential trigger for adverse events in left ventricular assist device application. Activation of coagulation showed a time-dependent increase, with no differences between both groups. Conclusions: This experimental study confirms the hypothesis that axial-flow pumps may induce stronger hemolysis compared to centrifugal pumps, coming along with larger amounts of circulating extracellular vesicles and a stronger PLT activation.Keywords
Funding Information
- Deutsche Forschungsgemeinschaft (LE 817/37-1, RO 2000/23-1, BR-5308/3-1)
This publication has 58 references indexed in Scilit:
- Axial and centrifugal continuous-flow rotary pumps: A translation from pump mechanics to clinical practiceThe Journal of Heart and Lung Transplantation, 2013
- A Quantitative Comparison of Mechanical Blood Damage Parameters in Rotary Ventricular Assist Devices: Shear Stress, Exposure Time and Hemolysis IndexJournal of Biomechanical Engineering, 2012
- Enhanced microparticles in ventricular assist device patients predict platelet, leukocyte and endothelial cell activationInteractive CardioVascular and Thoracic Surgery, 2010
- The effect of topography of polymer surfaces on platelet adhesionBiomaterials, 2010
- Hematological Problems During the Use of Cardiac Assist Devices: Clinical Experiences in JapanArtificial Organs, 2008
- Nano-scale superhydrophobicity: suppression of protein adsorption and promotion of flow-induced detachmentLab on a Chip, 2008
- A micro-fluidic study of whole blood behaviour on PMMA topographical nanostructuresJournal of Nanobiotechnology, 2008
- Microtextured Materials for Circulatory Support Devices: Preliminary StudiesASAIO Journal, 2006
- Properties of blood‐contacting surfaces of clinically implanted cardiac assist devices: Gene expression, matrix composition, and ultrastructural characterization of cellular liningsJournal of Cellular Biochemistry, 1995
- Dynamic fluctuations in blood of thrombin/antithrombin III complex (TAT)American Journal of Hematology, 1991