Effect of air leak on the performance of auto-PAP devices: a bench study
- 5 November 2005
- journal article
- Published by Springer Nature in Sleep and Breathing
- Vol. 9 (4) , 167-175
Abstract
Automatic positive airway pressure (auto-PAP) devices, used in the treatment of patients with obstructive sleep apnea (OSA), may not function optimally in the presence of an air leak. We set out to determine the magnitude of air leak that prevents auto-PAP devices from responding to respiratory events of OSA in a bench model. We simulated apnea, flow limitation, obstructive hypopnea, nonobstructive hypopnea, and snoring events of OSA with an artificial airway and a loudspeaker in a bench model connected to a commercially available auto-PAP device. Four auto-PAP devices were tested, but two of the “older-generation” devices (Tranquility and Virtuoso) did not respond to events of OSA that involved changes in flow contour; hence, we studied the effects of air leak and humidifier in the two “newer-generation” auto-PAP devices only (GoodKnight 418P, Autoset-T). When the air leak was progressively increased from baseline levels recommended by the manufacturer to levels seen clinically—5 to 7, 10, and 30 l/min—the GoodKnight 418P decreased pressure response by 56% (5.6±1.8 cm H2O, p=0.04). The pressure response of the Autoset-T, however, did not change from baseline during similar levels of air leak. The GoodKnight performed appropriately when the air leak was within 20 l/min, but the corresponding value for the Autoset was higher at 50 l/min. For both devices combined, air leak caused the pressure to drop between the device and the airway: 2.8±0.3 cm H2O at 30 l/min of air leak (p<0.001). Air leaks cause auto-PAP devices to underestimate the pressure required to treat events of OSA and to overestimate the pressure delivered at the upper airway. Physicians should be aware of performance limitations of auto-PAP devices in the presence of an air leak.Keywords
This publication has 26 references indexed in Scilit:
- How Are Automatic Positive Airway Pressure and Related Devices Prescribed by Sleep Physicians? A Web-Based SurveyJournal of Clinical Sleep Medicine, 2005
- Randomized Short-term Trial of Two AutoCPAP Devices versus Fixed Continuous Positive Airway Pressure for the Treatment of Sleep ApneaAmerican Journal of Respiratory and Critical Care Medicine, 2003
- Response of Automatic Continuous Positive Airway Pressure Devices to Different Sleep Breathing Patterns: A Bench StudyAmerican Journal of Respiratory and Critical Care Medicine, 2002
- Chemical Control Stability in Patients with Obstructive Sleep ApneaAmerican Journal of Respiratory and Critical Care Medicine, 2001
- Comparison of Manual and Automatic CPAP Titration in Patients With Sleep Apnea/Hypopnea SyndromeAmerican Journal of Respiratory and Critical Care Medicine, 1996
- Evaluation of an auto-nCPAP device based on snoring detectionEuropean Respiratory Journal, 1996
- Unattended Continuous Positive Airway Pressure Titration. Clinical Relevance and Cardiorespiratory Hazards of the MethodAmerican Journal of Respiratory and Critical Care Medicine, 1996
- Efficacy of Auto-CPAP in the Treatment of Obstructive Sleep Apnea/Hypopnea SyndromeAmerican Journal of Respiratory and Critical Care Medicine, 1996
- Indications and Standards for Use of Nasal Continuous Positive Airway Pressure (CPAP) in Sleep Apnea Syndromes. American Thoracic Society. Official Statement Adopted March 1944American Journal of Respiratory and Critical Care Medicine, 1994
- REVERSAL OF OBSTRUCTIVE SLEEP APNOEA BY CONTINUOUS POSITIVE AIRWAY PRESSURE APPLIED THROUGH THE NARESThe Lancet, 1981