Robust time-shifted spoke pulse design in the presence of large B0 variations with simultaneous reduction of through-plane dephasing, B1+ effects, and the specific absorption rate using parallel transmission
- 7 October 2015
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 76 (2) , 540-554
Abstract
Purpose To design parallel transmission spokes pulses with time‐shifted profiles for joint mitigation of intensity variations due to B1+ effects, signal loss due to through‐plane dephasing, and the specific absorption rate (SAR) at 7T. Methods We derived a slice‐averaged small tip angle (SA‐STA) approximation of the magnetization signal at echo time that depends on the B1+ transmit profiles, the through‐slice B0 gradient and the amplitude and time‐shifts of the spoke waveforms. We minimize a magnitude least‐squares objective based on this signal equation using a fast interior‐point approach with analytical expressions of the Jacobian and Hessian. Results Our algorithm runs in less than three minutes for the design of two‐spoke pulses subject to hundreds of local SAR constraints. On a B0/B1+ head phantom, joint optimization of the channel‐dependent time‐shifts and spoke amplitudes allowed signal recovery in high‐B0 regions at no increase of SAR. Although the method creates uniform magnetization profiles (ie, uniform intensity), the flip angle varies across the image, which makes it ill‐suited to T1‐weighted applications. Conclusions The SA‐STA approach presented in this study is best suited to T2*‐weighted applications with long echo times that require signal recovery around high B0 regions. Magn Reson Med 76:540–554, 2016.Keywords
Funding Information
- National Institutes of Health (R01-EB006847-08, P41-EB015896-17)
This publication has 32 references indexed in Scilit:
- Multi‐atlas and label fusion approach for patient‐specific MRI based skull estimationMagnetic Resonance in Medicine, 2015
- Automated slice‐specific simultaneous z‐shim method for reducing B1 inhomogeneity and susceptibility‐induced signal loss with parallel transmission at 3TMagnetic Resonance in Medicine, 2014
- A 16‐channel dual‐row transmit array in combination with a 31‐element receive array for human brain imaging at 9.4 TMagnetic Resonance in Medicine, 2013
- Spectral decomposition of susceptibility artifacts for spectral‐spatial radiofrequency pulse designMagnetic Resonance in Medicine, 2012
- Fast MRI coil analysis based on 3-D electromagnetic and RF circuit co-simulationJournal of Magnetic Resonance, 2009
- Reducing susceptibility artifacts in fMRI using volume‐selective z‐shim compensationMagnetic Resonance in Medicine, 2007
- Spatial domain method for the design of RF pulses in multicoil parallel excitationMagnetic Resonance in Medicine, 2006
- EPI image reconstruction with correction of distortion and signal lossesJournal of Magnetic Resonance Imaging, 2006
- Compensation of Susceptibility-Induced BOLD Sensitivity Losses in Echo-Planar fMRI ImagingNeuroImage, 2002
- MR susceptibility misregistration correctionIEEE Transactions on Medical Imaging, 1993