How is photon head scatter in a linear accelerator related to the concept of a virtual source?
Authors:
- 1 August 1993
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 38 (8) , 1173-1180
Abstract
The virtual source in a photon linear accelerator is a familiar concept which arises from head scatter, and which accounts for small but detectable deviations of photon energy fluence from the inverse-square law. The authors present a simple method for calculating the position of the virtual source using the geometry of the flattening filter, which acts as the main photon scatterer. The authors also show that, for unwedged beams, the position of the virtual source is relatively insensitive to collimator size. This means that the same head-scatter factor which is measured at the isocentre can also be used at other distances from the source. However, wedges may scatter so much radiation that the position of the virtual source depends upon collimator size.Keywords
This publication has 8 references indexed in Scilit:
- Head-scatter factors and effective x-ray source positions in a 25-MV linear acceleratorMedical Physics, 1992
- Application of the convolution method for calculation of output factors for therapy photon beamsMedical Physics, 1992
- The determination of phantom and collimator scatter components of the output of megavoltage photon beams: measurement of the collimator scatter part with a beam-coaxial narrow cylindrical phantomRadiotherapy and Oncology, 1991
- Separation of primary and scatter components of measured photon beam dataPhysics in Medicine & Biology, 1989
- Output factor constituents of a high‐energy photon beamMedical Physics, 1988
- Head scatter data for several linear accelerators (4-18 MV)Medical Physics, 1986
- Characteristics of an 18 MV photon beam from a Therac 20 Medical Linear AcceleratorMedical Physics, 1981
- The Extension of the Concept of Tissue-Air Ratios (TAR) to High-Energy X-Ray BeamsRadiology, 1970