Effects of sterilization on the Tekscan digital pressure sensor

Authors:
  • VHValerie S Harder
    Valerie S Harder
  • ELEugene P Lautenschlager
    Eugene P Lautenschlager
  • JKJames C Kudrna
    James C Kudrna

Abstract
Investigations into the effects of sterilization on a new biomechanical pressure sensor are necessary before contemplating in vivo use. Ten, designated Experimental, "K-Scan" digital pressure sensor arrays were sterilized with ethylene oxide gas (EtO), and their ability to accurately and reproducibly measure an applied load of 2225 N (500 lb) was assessed. Simultaneously, 10 un-sterilized sensor arrays, designated Control, were assessed. Each array was loaded 10 times inside a two-dimensional curved surface, and all arrays exhibited high reproducibility (coefficients of variation0.05, beta0.05, beta<0.05). Results suggest that, following EtO sterilization, accurate and reproducible pressure measurements can be obtained from K-Scan sensors when calibration is performed at time of use.