Radiation Dose at Coronary CT Angiography: Second-Generation Dual-Source CT Versus Single-Source 64-MDCT and First-Generation Dual-Source CT

Authors:

Abstract
OBJECTIVE. The purpose of this study was to assess the radiation doses of different coronary CTA (CTA) protocols: second-generation dual-source 128-MDCT, first-generation dual-source 64-MDCT, and single-source 64-MDCT.MATERIALS AND METHODS. Thermoluminescent dosimetry was used to determine scanner-specific dose coefficients for standard coronary CTA of an anthropomorphic phantom. These coefficients were used to estimate the effective doses (EDs) of retrospectively gated, prospectively triggered, and prospectively triggered high pitch coronary CTA performed at 100 and 120 kV. The coronary CTA protocols used in imaging of 43 patients undergoing dual-source 128-MDCT were analyzed for ED, image quality, and signal-to-noise ratio.RESULTS. Regardless of coronary CTA protocol and CT system, imaging at 100 kV lowered the ED 40–50%. In retrospectively gated 120-kV coronary CTA, the ED ranged from 5.7 to 10.7 mSv and was approximately 50% lower with single-source 64-MDCT than with either DSCT protocol. In prospecti...

This publication has 31 references indexed in Scilit: