Registration uncertainties between 3D cone beam computed tomography and different reference CT datasets in lung stereotactic body radiation therapy
Open Access
- 26 October 2016
- journal article
- research article
- Published by Springer Nature in Radiation Oncology
- Vol. 11 (1) , 142
Abstract
The aim of this study was to analyze differences in couch shifts (setup errors) resulting from image registration of different CT datasets with free breathing cone beam CTs (FB-CBCT). As well automatic as manual image registrations were performed and registration results were correlated to tumor characteristics. FB-CBCT image registration was performed for 49 patients with lung lesions using slow planning CT (PCT), average intensity projection (AIP), maximum intensity projection (MIP) and mid-ventilation CTs (MidV) as reference images. Both, automatic and manual image registrations were applied. Shift differences were evaluated between the registered CT datasets for automatic and manual registration, respectively. Furthermore, differences between automatic and manual registration were analyzed for the same CT datasets. The registration results were statistically analyzed and correlated to tumor characteristics (3D tumor motion, tumor volume, superior-inferior (SI) distance, tumor environment). Median 3D shift differences over all patients were between 0.5 mm (AIPvsMIP) and 1.9 mm (MIPvsPCT and MidVvsPCT) for the automatic registration and between 1.8 mm (AIPvsPCT) and 2.8 mm (MIPvsPCT and MidVvsPCT) for the manual registration. For some patients, large shift differences (>5.0 mm) were found (maximum 10.5 mm, automatic registration). Comparing automatic vs manual registrations for the same reference CTs, ∆AIP achieved the smallest (1.1 mm) and ∆MIP the largest (1.9 mm) median 3D shift differences. The standard deviation (variability) for the 3D shift differences was also the smallest for ∆AIP (1.1 mm). Significant correlations (p < 0.01) between 3D shift difference and 3D tumor motion (AIPvsMIP, MIPvsMidV) and SI distance (AIPvsMIP) (automatic) and also for 3D tumor motion (∆PCT, ∆MidV; automatic vs manual) were found. Using different CT datasets for image registration with FB-CBCTs can result in different 3D couch shifts. Manual registrations achieved partly different 3D shifts than automatic registrations. AIP CTs yielded the smallest shift differences and might be the most appropriate CT dataset for registration with 3D FB-CBCTs.Keywords
This publication has 42 references indexed in Scilit:
- Accuracy and inter-observer variability of 3D versus 4D cone-beam CT based image-guidance in SBRT for lung tumorsRadiation Oncology, 2012
- Quantification of the Variability of Diaphragm Motion and Implications for Treatment Margin ConstructionInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012
- Comparison of setup error using different reference images: a phantom and lung cancer patients studyMedical Dosimetry, 2011
- The potential for undertaking slow CT using a modern CT scannerBritish Journal of Radiology, 2010
- Potential of image-guidance, gating and real-time tracking to improve accuracy in pulmonary stereotactic body radiotherapyRadiotherapy and Oncology, 2009
- Frameless Stereotactic Body Radiotherapy for Lung Cancer Using Four-Dimensional Cone Beam CT GuidanceInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Use of Maximum Intensity Projections (MIPs) for Target Outlining in 4DCT Radiotherapy PlanningJournal of Thoracic Oncology, 2008
- On-Line Target Position Localization in the Presence of Respiration: A Comparison of two MethodsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007
- Cone-beam CT based image-guidance for extracranial stereotactic radiotherapy of intrapulmonary tumorsActa Oncologica, 2006
- Measurement of lung tumor motion using respiration-correlated CTInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2004