Characterizing the structural properties and porosity of mid-urethral slings with varied manufacturing techniques
Authors:
Open Access
- 7 August 2025
- journal article
- research article
- Published by Frontiers Media SA in Frontiers in Bioengineering and Biotechnology
- Vol. 13, 1543808
Abstract
Introduction: Implantation of mid-urethral slings (MUSs) is a safe and effective approach for the surgical repair of stress urinary incontinence. However, concerns regarding the deformability of the prototype MUS, mechanical cut Gynecare TVT prompted manufacturers to use techniques like laser cutting, heat-sealing, and the inclusion of an interwoven stabilizing suture to decrease deformation with loading. We hypothesized that a laser cut or heat-sealed MUS would be stiffer but deform less, and experience less permanent elongation as compared to a mechanical cut MUS. Additionally, the inclusion of a stabilizing suture would minimize the loss of porosity.Methods: Uniaxial tensile testing to failure and cyclic loading was performed to analyze the structural properties and permanent elongation, respectively, of commercially available MUSs Gynecare TVT (mechanical cut), Gynecare TVT Exact (laser cut), ArcTV (laser cut, with and without the stabilizing suture), and Desara Blue (heat-sealed). A custom Mathematica code was used to quantify the porosity of the MUSs following sequential uniaxial loading from 0 N to 10 N.Results: Desara Blue was significantly stiffer (p-values < 0.05), elongated less at failure (p = 0.002), and experienced less permanent elongation in response to cyclic loading (p-values = 0.001) relative to Gynecare TVT. Similarly, permanent elongation was significantly less (p-values = 0.004) and the stiffness was higher (p = 0.004) for Gynecare TVT Exact as compared to Gynecare TVT. Very little differences in stiffness and no differences in relative elongation at failure nor permanent elongation were observed between ArcTV (without the interwoven suture) and Gynecare TVT (p-values > 0.05). The porosity of all MUSs significantly decreased with loading (p-values < 0.001); except for ArcTV with the stabilizing suture which showed the least amount of deformation (i.e., percent change in porosity decreasing by only 14%, p < 0.001, at 10 N).Discussion: Overall, heat-sealing decreased deformability at the cost of markedly increasing device stiffness to a point which likely outweighs benefits, and risks increased complications. Laser cutting had different effects on the behavior of TVT Exact and ArcTV suggesting manufacturer technical differences, but overall reduced deformation without a substantial impact on stiffness. An interwoven stabilizing suture minimized the loss of porosity which translates clinically to less deformation and mechanistically to reduced mesh complications.Keywords
This publication has 51 references indexed in Scilit:
- Tensile properties of five commonly used mid-urethral slings relative to the TVT™International Urogynecology Journal, 2008
- The influence of porosity on the integration histology of two polypropylene meshes for the treatment of abdominal wall defects in dogsHernia, 2007
- Complications of tension‐free vaginal tape surgery: a multi‐institutional reviewBJU International, 2004
- The prevalence of urinary incontinence in women in four European countriesBJU International, 2004
- The definition, prevalence, and risk factors for stress urinary incontinence.2004
- Impact of Polymer Pore Size on the Interface Scar Formation in a Rat ModelJournal of Surgical Research, 2002
- The influence of differing pore sizes on the biocompatibility of two polypropylene meshes in the repair of abdominal defectsHernia, 2001
- Differential effects of cough, valsalva, and continence status on vesical neck movementObstetrics & Gynecology, 2000
- An ambulatory surgical procedure under local anesthesia for treatment of female urinary incontinenceInternational Urogynecology Journal, 1996
- Effects of Stress Shielding on the Mechanical Properties of Rabbit Patellar TendonJournal of Biomechanical Engineering, 1993