Determination of Thyroxine Using Tris(2,2′‐Bipyridyl)Ruthenium(III)‐NADH Enhanced Electrochemiluminescence Detection

Authors:

Abstract
A low cost, single channel portable flow injection electrogenerated chemiluminescence (ECL) system is reported for the determination of l‐thyroxine based on its enhancement effect on Ru(bpy)3 3+‐NADH‐ECL intensity. The calibration graph was linear over the range 5.0×10−8–1.0×10−6 mol l−1 (R 2=0.9981) with relative standard deviations (RSD, n=4) in the range 0.11–1.5%. The limit of detection (3σ blank) was 5.0×10−8 mol l−1 with a sample throughput of 60 h−1. The method was applied to pharmaceutical preparations and the results obtained (51.2±1.23–52±1.14 µg l‐thyroxine per tablet) were in good agreement with the amount labeled (50 µg l‐thyroxine per tablet, sodium salt) and RIA method (50.2±1.25 µg l‐thyroxine per tablet) as the reference method. NADH calibration graph (in the range 5.0×10−8–1.8×10−6 mol l−1, R 2=0.9989) was also established with RSD (n=4) in the range 0.25–2.1% with a limit of detection (3σ blank) 5.0×10−8 mol l−1.