Microfabricated On-Line Sensor for Continuous Monitoring of l-Glutamate
Authors:
- 2 October 1998
- journal article
- Published by Springer Nature in Analytical Sciences
- Vol. 14 (5) , 947-953
Abstract
An on-line sensor was developed by using a micro-machining technique for the continuous measurement of extracellular glutamate. The sensor was made by bonding two glass plates together. One of the plates has a rectangular flow channel which is connected to both a sampling and an outlet capillary. The other has a carbon film-based electrochemical cell, which consists of an enzyme-modified electrode surrounded by polymer film to form a thin-layer-channel flow cell. The sensor has a high analyte conversion efficiency due to the cell structure. As a result, we obtained a low detection limit of 44 nM (S/N=3). The cathodic current started to increase about 10 s after sample introduction at a flow rate of 2 μl/min due to its small inner volume. Using the sensor, we can monitor the extracellular L-glutamate increase caused by stimulating it with KCl and γ-aminobutyric acid (GABA) by locating the sampling capillary very close to the cell.Keywords
This publication has 40 references indexed in Scilit:
- Design and development of a miniaturised total chemical analysis system for on-line lactate and glucose monitoring in biological samplesAnalytica Chimica Acta, 1997
- Continuous monitoring of L-glutamate released from cultured nerve cells by an online sensor coupled with micro-capillary samplingBiosensors and Bioelectronics, 1997
- Amperometric Biosensors for On-Line Monitoring of Extracellular Glucose and Glutamate in the BrainAnalytical Letters, 1994
- Dehydrogenase-modified carbon-fiber microelectrodes for the measurement of neurotransmitter dynamics. 1. NADH voltammetryAnalytical Chemistry, 1993
- Hydrogen peroxide and .beta.-nicotinamide adenine dinucleotide sensing amperometric electrodes based on electrical connection of horseradish peroxidase redox centers to electrodes through a three-dimensional electron relaying polymer networkAnalytical Chemistry, 1992
- The dialysis electrode—a new method for in vivo monitoringJournal of the Chemical Society, Chemical Communications, 1992
- Kinetic-potentiometric determination of monosodium glutamate in soups and soup bases and of glutamic dehydrogenaseThe Analyst, 1987
- Enzyme Electrodes For L-Glutamate Using Chemical Redox Mediators and Enzymatic Substrate AmplificationAnalytical Letters, 1986
- Elevation of the Extracellular Concentrations of Glutamate and Aspartate in Rat Hippocampus During Transient Cerebral Ischemia Monitored by Intracerebral MicrodialysisJournal of Neurochemistry, 1984
- A miniature flow-through cell with a four-function chemfet integrated circuit for simultaneous measurements of potassium, hydrogen, calcium and sodium ionsAnalytica Chimica Acta, 1984