Determination of the reference air kerma rate for brachytherapy sources and the related uncertainty
Authors:
- Eduard van DijkEDEduard van Dijk
- Inger‐Karine K. Kolkman‐DeurlooIKInger‐Karine K. Kolkman‐Deurloo
- Patricia M. G. DamenPDPatricia M. G. Damen
- 21 September 2004
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 31 (10) , 2826-2833
Abstract
Different methods exist to determine the air kerma calibration factor of an ionization chamber for the spectrum of a high‐dose‐rate (HDR) or pulsed‐dose‐rate (PDR) source. An analysis of two methods to obtain such a calibration factor was performed: (i) the method recommended by [Goetsch et al., Med. Phys. 18, 462–467 (1991)] and (ii) the method employed by the Dutch national standards institute NMi [Petersen et al., Report S‐EI‐94.01 (NMi, Delft, The Netherlands, 1994)]. This analysis showed a systematic difference on the order of 1% in the determination of the strength of HDR and PDR sources depending on the method used for determining the air kerma calibration factor. The definitive significance of the difference between these methods can only be addressed after performing an accurate analysis of the associated uncertainties. For an NE 2561 (or equivalent) ionization chamber and an in‐air jig, a typical uncertainty budget of 0.94% was found with the NMi method. The largest contribution in the type‐B uncertainty is the uncertainty in the air kerma calibration factor for isotope , , as determined by the primary or secondary standards laboratories. This uncertainty is dominated by the uncertainties in the physical constants for the average mass‐energy absorption coefficient ratio and the stopping power ratios. This means that it is not foreseeable that the standards laboratories can decrease the uncertainty in the air kerma calibration factors for ionization chambers in the short term. When the results of the determination of the reference air kerma rates in, e.g., different institutes are compared, the uncertainties in the physical constants are the same. To compare the applied techniques, the ratio of the results can be judged by leaving out the uncertainties due to these physical constants. In that case an uncertainty budget of 0.40% (coverage ) should be taken into account. Due to the differences in approach between the method used by NMi and the method recommended by Goetsch et al., an extra type‐B uncertainty of 0.9% has to be taken into account when the method of Goetsch et al. is applied. Compared to the uncertainty of 1% found for the air calibration of , the difference of 0.9% found is significant.Keywords
This publication has 13 references indexed in Scilit:
- Calibration of new high dose rate sourcesMedical Physics, 2002
- Experimental derivation of wall correction factors for ionization chambers used in high dose rate source calibrationMedical Physics, 2001
- Monte Carlo study of correction factors for Spencer–Attix cavity theory at photon energies at or above 100 keVMedical Physics, 2000
- Code of practice for brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 56Medical Physics, 1997
- Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43Medical Physics, 1995
- Calibration of low activity192Ir brachytherapy sources in terms of reference air kerma rate with large volume spherical ionization chambersPhysics in Medicine & Biology, 1992
- A new re-entrant ionization chamber for the calibration of iridium-192 high dose rate sourcesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1992
- Calibration of 192Ir high‐dose‐rate afterloading systemsMedical Physics, 1991
- Wall attenuation and scatter corrections for ion chambers: measurements versus calculationsPhysics in Medicine & Biology, 1990
- On the technique of extrapolation to obtain wall correction factors for ion chambers irradiated by photon beamsMedical Physics, 1990