Good mid-term results with the trident peripheral self-locking cup: a clinical evaluation and migration measurement with EBRA
Open Access
- 9 November 2020
- journal article
- research article
- Published by Springer Nature in Archives of orthopaedic and trauma surgery
- Vol. 141 (2) , 327-332
Abstract
Introduction: The most common cause of failure in total hip arthroplasty (THA) is aseptic loosening. Uncemented cup migration analysis by means of Einzel–Bild–Roentgen–Analyse (EBRA) has shown to be a good predictive indicator for early implant failure if the cup migrates more than 2 mm within 4 years after surgery. In this study, we performed a migration analysis of an uncemented peripheral self-locking (PSL) press-fit cup after 4 years follow-up. Materials and methods: We retrospectively reviewed all patients who received a trident PSL press-fit cup at our department between 2004 and 2017. A total of 636 patients were identified. As inclusion criteria for radiological analysis, a minimum follow-up of 2 years was defined. We reviewed medical histories and performed radiological analysis using EBRA software. EBRA measurements and statistical investigations were performed by two independent investigators. Results: A total of 149 cups in 146 patients (female 82; male 64) met our inclusion criteria. Mean age at surgery was 65 years (33–89). We found a significant improvement in the WOMAC score pre- to postoperative (p < 0.0001). EBRA migration analysis showed a mean total migration of 0.6 mm (0.0–8.2) over our follow-up period of 4 years. Of the investigated cups, 69.8% showed a migration rate smaller than 2 mm in the investigated follow-up. Conclusion: The acetabular cup used in our study provides low migration at final follow-up. Therefore, a good long-term outcome can be expected for the PSL cup. Trial registration: Trial registration number is 20181024-1875 and date of registration is 2018-10-24.Keywords
Funding Information
- University of Innsbruck and Medical University of Innsbruck
This publication has 23 references indexed in Scilit:
- Five-year Results of the Uncemented RM Pressfit Cup Clinical Evaluation and Migration Measurements by EBRAThe Journal of Arthroplasty, 2013
- Two-Year Radiologic Assessment of the Trident Peripheral Self-Locking Cup Using EbraHIP International, 2012
- Migration of the Duraloc cup after 5 yearsInternational Orthopaedics, 2007
- Patient-related predictors of implant failure after primary total hip replacement in the initial, short- and long-termsThe Journal of Bone and Joint Surgery. British volume, 2006
- Modes of failure of a threaded acetabular cupInternational Orthopaedics, 2006
- Bipolar revision arthroplasty for failed threaded acetabular components: radiographic evaluation of cup migrationJournal of Orthopaedic Science, 2002
- Accuracy of EBRA-FCA in the measurement of migration of femoral components of total hip replacementThe Journal of Bone and Joint Surgery. British volume, 1999
- EBRA improves the accuracy of radiographic analysis of acetabular cup migrationActa Orthopaedica, 1998
- EBRA: A method to measure migration of acetabular componentsJournal of Biomechanics, 1995
- Measurement accuracy in acetabular cup migrationThe Journal of Arthroplasty, 1992