Calibration of a surgical microscope with automated zoom lenses using an active optical tracker
- 14 February 2008
- journal article
- research article
- Published by Wiley in The International Journal of Medical Robotics and Computer Assisted Surgery
- Vol. 4 (1) , 87-93
Abstract
Background: In this paper, we present a new method for the calibration of a microscope and its registration using an active optical tracker.Methods: Practically, both operations are done simultaneously by moving an active optical marker within the field of view of the two devices. The IR LEDs composing the marker are first segmented from the microscope images. By knowing their corresponding three‐dimensional (3D) position in the optical tracker reference system, it is possible to find the transformation matrix between the referential of the two devices. Registration and calibration parameters can be extracted directly from that transformation. In addition, since the zoom and focus can be modified by the surgeon during the operation, we propose a spline based method to update the camera model to the new setup.Results: The proposed technique is currently being used in an augmented reality system for image‐guided surgery in the fields of ear, nose and throat (ENT) and craniomaxillofacial surgeries.Conclusions: The results have proved to be accurate and the technique is a fast, dynamic and reliable way to calibrate and register the two devices in an OR environment. Copyright © 2008 John Wiley & Sons, Ltd.Keywords
This publication has 12 references indexed in Scilit:
- A fully automated calibration method for an optical see-through head-mounted operating microscope with variable zoom and focusIEEE Transactions on Medical Imaging, 2005
- A flexible new technique for camera calibrationIEEE Transactions on Pattern Analysis and Machine Intelligence, 2000
- Design and evaluation of a system for microscope-assisted guided interventions (MAGI)IEEE Transactions on Medical Imaging, 2000
- Geometric camera calibration using circular control pointsIEEE Transactions on Pattern Analysis and Machine Intelligence, 2000
- Predicting error in rigid-body point-based registrationIEEE Transactions on Medical Imaging, 1998
- Computer Navigational Microscope for Minimally Invasive Neurosurgerymin - Minimally Invasive Neurosurgery, 1997
- Image guided microscopic surgery system using mutual-information based registration1996
- Modeling and calibration of automated zoom lenses1994
- Principal warps: thin-plate splines and the decomposition of deformationsIEEE Transactions on Pattern Analysis and Machine Intelligence, 1989
- A least-squares algorithm for the equiform transformation from spatial marker co-ordinatesJournal of Biomechanics, 1988