A novel nonlinear afterload for ex vivo heart evaluation: Porcine experimental results
Authors:
Open Access
- 20 May 2022
- journal article
- research article
- Published by Wiley in Artificial Organs
- Vol. 46 (9) , 1794-1803
Abstract
Background: Existing working heart models for ex vivo functional evaluation of donor hearts often use cardiac afterloads made up of discrete resistive and compliant elements. This approach limits the practicality of independently controlling systolic and diastolic aortic pressure to safely test the heart under multiple loading conditions. We present and investigate a novel afterload concept designed to enable such control.Methods: Six ∼70 kg pig hearts were evaluated in vivo, then ex vivo in left‐ventricular working mode using the presented afterload. Both in vivo and ex vivo, the hearts were evaluated at two exertion levels: at rest and following a 20 μg adrenaline bolus, while measuring aortic pressure and flow, left ventricular pressure and volume, and left atrial pressure.Results: The afterload gave aortic pressure waveforms that matched the general shape of the in vivo measurements. A wide range of physiological systolic pressures (93 to 160 mm Hg) and diastolic pressures (73 to 113 mm Hg) were generated by the afterload.Conclusions: With the presented afterload concept, multiple physiological loading conditions could be tested ex vivo, and compared with the corresponding in vivo data. An additional control loop from the set pressure limits to the measured systolic and diastolic aortic pressure is proposed to address discrepancies observed between the set limits and the measured pressures.Keywords
Funding Information
- Hans-Gabriel och Alice Trolle-Wachtmeisters stiftelse för medicinsk forskning
- Stiftelsen för Strategisk Forskning (SM21‐0037)
- Vetenskapsrådet (2017‐04989)
This publication has 21 references indexed in Scilit:
- Large-Animal Biventricular Working Heart Perfusion System with Low Priming Volume—Comparison between in vivo and ex vivo Cardiac FunctionThe Thoracic and Cardiovascular Surgeon, 2016
- Safe orthotopic transplantation of hearts harvested 24 hours after brain death and preserved for 24 hoursScandinavian Cardiovascular Journal, 2016
- Assessment of donor heart viability during ex vivo heart perfusionCanadian Journal of Physiology and Pharmacology, 2015
- Ex-vivo perfusion of donor hearts for human heart transplantation (PROCEED II): a prospective, open-label, multicentre, randomised non-inferiority trialThe Lancet, 2015
- Successful Heart Transplant after Ten Hours Out-of-body Time using the TransMedics Organ Care SystemHeart, Lung and Circulation, 2015
- Ex vivo heart perfusion after cardiocirculatory death; a porcine modelJournal of Surgical Research, 2015
- Pharmacological normalization of circulation after acute brain deathActa Anaesthesiologica Scandinavica, 2012
- An Ex Vivo Platform to Simulate Cardiac Physiology: A New Dimension for Therapy Development and AssessmentThe International Journal of Artificial Organs, 2011
- Development of a Physical Windkessel Module to Re-Create In Vivo Vascular Flow Impedance for In Vitro ExperimentsCardiovascular Engineering and Technology, 2010
- The arterial WindkesselMedical & Biological Engineering & Computing, 2008